VentrAssist Left Ventricular Assist Systems Long-Term In Vitro Reliability

Alex Stonehouse

Alex Stonehouse

Ventracor Limited, Chatswood, NSW, Australia

Search for more papers by this author
Martin Cook

Martin Cook

Ventracor Limited, Chatswood, NSW, Australia

Search for more papers by this author
John Woodard

John Woodard

Ventracor Limited, Chatswood, NSW, Australia

Search for more papers by this author
First published: 20 October 2009
Citations: 3
Dr. John Woodard, 126 Greville Street, Chatswood, NSW, Australia. E-mail: [email protected]

Abstract

Left ventricular assist systems (LVADs) are typically tested for long-term reliability by using a simulated human cardiovascular system. In this study, 13 VentrAssist (Ventracor Ltd, Chatswood, NSW, Australia) systems are tested in mock loops with the aim of demonstrating in vitro continuous operational reliability of the pump subsystem greater than 80% (60% confidence) over 1 year. As of July 9, 2007, the total time on test of all 13 mock loop pumps is 29.36 years. In vitro reliability of the VentrAssist LVAD blood pump is 96.7% (60% confidence) at 1 year. No failures have been seen over 4.5 years of test, and testing is ongoing.