Volume 56, Issue 1 pp. 224-229
Note

Enhancing the precision of quantitative water content mapping by optimizing sequence parameters

H. Neeb

H. Neeb

Institut für Medizin, Forschungszentrum Jülich GmbH, Jülich, Germany

Search for more papers by this author
N.J. Shah

Corresponding Author

N.J. Shah

Institut für Medizin, Forschungszentrum Jülich GmbH, Jülich, Germany

Brain Imaging Center West, Jülich, Germany

Institute of Physics, University of Dortmund, Dortmund, Germany

Institut für Medizin, Forschungszentrum Jülich GmbH, 52425 Jülich, Germany===Search for more papers by this author
First published: 26 May 2006
Citations: 12

Abstract

This article describes a methodology for determining the precision of a recently published method for quantitative water content mapping using MRI. A framework for calculating the random error component as a function of sequence parameters is provided and predictions are compared with experimental results obtained in a phantom. The experimental and theoretical results are in good agreement and demonstrate that the precision of quantitative water content mapping can be significantly increased by a proper and objective choice of acquisition parameters. Magn Reson Med, 2006. © 2006 Wiley-Liss, Inc.