Volume 34, Issue 1 pp. 22-30
Original Research

Intracranial arterial wall imaging using three-dimensional high isotropic resolution black blood MRI at 3.0 Tesla

Ye Qiao PhD

Ye Qiao PhD

The Russell H. Morgan Department of Radiology and Radiological Sciences, The Johns Hopkins Hospital, Baltimore, Maryland, USA

Search for more papers by this author
David A. Steinman PhD

David A. Steinman PhD

Biomedical Simulation Laboratory, Department of Mechanical and Industrial Engineering, University of Toronto, Toronto, Ontario, Canada

Search for more papers by this author
Qin Qin PhD

Qin Qin PhD

The Russell H. Morgan Department of Radiology and Radiological Sciences, The Johns Hopkins Hospital, Baltimore, Maryland, USA

F.M. Kirby Research Center for Functional Brain Imaging, Kennedy Krieger Institute, Baltimore, Maryland, USA

Search for more papers by this author
Maryam Etesami MD

Maryam Etesami MD

The Russell H. Morgan Department of Radiology and Radiological Sciences, The Johns Hopkins Hospital, Baltimore, Maryland, USA

Search for more papers by this author
Michael Schär PhD

Michael Schär PhD

The Russell H. Morgan Department of Radiology and Radiological Sciences, The Johns Hopkins Hospital, Baltimore, Maryland, USA

Philips Healthcare, Cleveland, Ohio, USA

Search for more papers by this author
Brad C. Astor PhD

Brad C. Astor PhD

Department of Epidemiology, The Johns Hopkins Bloomberg School of Public Health, Baltimore, Maryland, USA

Department of Medicine, The Johns Hopkins University School of Medicine, Baltimore, Maryland, USA

Search for more papers by this author
Bruce A. Wasserman MD

Corresponding Author

Bruce A. Wasserman MD

The Russell H. Morgan Department of Radiology and Radiological Sciences, The Johns Hopkins Hospital, Baltimore, Maryland, USA

Johns Hopkins Hospital, 367 East Park Building, 600 North Wolfe Street, Baltimore, MD 21287Search for more papers by this author
First published: 22 June 2011
Citations: 233

Abstract

Purpose:

To develop a high isotropic-resolution sequence to evaluate intracranial vessels at 3.0 Tesla (T).

Materials and Methods:

Thirteen healthy volunteers and 4 patients with intracranial stenosis were imaged at 3.0T using 0.5-mm isotropic-resolution three-dimensional (3D) Volumetric ISotropic TSE Acquisition (VISTA; TSE, turbo spin echo), with conventional 2D-TSE for comparison. VISTA was repeated for 6 volunteers and 4 patients at 0.4-mm isotropic-resolution to explore the trade-off between SNR and voxel volume. Wall signal-to-noise-ratio (SNRwall), wall-lumen contrast-to-noise-ratio (CNRwall-lumen), lumen area (LA), wall area (WA), mean wall thickness (MWT), and maximum wall thickness (maxWT) were compared between 3D-VISTA and 2D-TSE sequences, as well as 3D images acquired at both resolutions. Reliability was assessed by intraclass correlations (ICC).

Results:

Compared with 2D-TSE measurements, 3D-VISTA provided 58% and 74% improvement in SNRwall and CNRwall-lumen, respectively. LA, WA, MWT and maxWT from 3D and 2D techniques highly correlated (ICCs of 0.96, 0.95, 0.96, and 0.91, respectively). CNRwall-lumen using 0.4-mm resolution VISTA decreased by 27%, compared with 0.5-mm VISTA but with reduced partial-volume-based overestimation of wall thickness. Reliability for 3D measurements was good to excellent.

Conclusion:

The 3D-VISTA provides SNR-efficient, highly reliable measurements of intracranial vessels at high isotropic-resolution, enabling broad coverage in a clinically acceptable time. J. Magn. Reson. Imaging 2011;. © 2011 Wiley-Liss, Inc.

The full text of this article hosted at iucr.org is unavailable due to technical difficulties.