Transradial coronary brachytherapy with the Novoste beta-rail system
Abstract
We report our initial experience in 10 consecutive patients who underwent transradial coronary brachytherapy for in-stent restenosis using a 90Sr/Y source and the Novoste Beta-Rail system. In all patients, procedures were successfully completed using a right transradial approach. We performed the procedures with the Beta-Rail catheter using 7 Fr (Zuma II, Medtronic, MN; n = 5) or 8 Fr (Cordis, Miami, FL; n = 5) guiding catheters. All lesions were successfully dilated and no additional stent was inserted. We used a 40 mm source (n = 3) or a 60 mm source (n = 7) with manual stepping in four cases. In three cases, we did one stepping, and in one case, we did three steppings. The mean dwell time was 195 ± 44 sec. The mean delivered dose was 23 ± 3 Gy at 2 mm distance from the source. No radiation treatment was interrupted. Mean fluoroscopy time was 26 ± 13 min. Procedural success was achieved in all patients. Three patients had mild CK elevations (< 3 times upper normal limit). All patients were pretreated with clopidogrel (300 mg) and combined treatment with aspirin + clopidogrel is to be continued for at least 1 year. Clinical follow-up up to 3 months has not yielded any complication and all patients have remained free from angina. Cathet Cardiovasc Intervent 2002;55:362–366. © 2002 Wiley-Liss, Inc.