research papers

Correction of the X-ray wavefront from compound refractive lenses using 3D printed refractive structures
aDiamond Light Source, Harwell Science and Innovation Campus, Didcot, Oxon OX11 0DE,
United Kingdom, bSynchrotron Utilisation Section, Raja Ramanna Centre for Advanced Technology, Indore
452012, India, cDepartment of Chemistry, Lancaster University, Lancaster LA1 4YB, United Kingdom,
and dDepartment of Engineering Science, Oxford University, Parks Road, Oxford OX1 3PJ,
United Kingdom
*Correspondence e-mail: [email protected]
A refractive phase corrector optics is proposed for the compensation of fabrication error of X-ray optical elements. Here, at-wavelength wavefront measurements of the focused X-ray beam by knife-edge imaging technique, the design of a three-dimensional corrector plate, its fabrication by 3D printing, and use of a corrector to compensate for X-ray lens figure errors are presented. A rotationally invariant corrector was manufactured in the polymer IP-STM using additive manufacturing based on the two-photon polymerization technique. The fabricated corrector was characterized at the B16 Test beamline, Diamond Light Source, UK, showing a reduction in r.m.s. wavefront error of a Be compound refractive Lens (CRL) by a factor of six. The r.m.s. wavefront error is a figure of merit for the wavefront quality but, for X-ray lenses, with significant X-ray absorption, a form of the r.m.s. error with weighting proportional to the transmitted has been proposed. The knife-edge imaging wavefront-sensing technique was adapted to measure rotationally variant wavefront errors from two different sets of Be CRL consisting of 98 and 24 lenses. The optical aberrations were then quantified using a Zernike polynomial expansion of the 2D wavefront error. The compensation by a rotationally invariant corrector plate was partial as the Be CRL wavefront error distribution was found to vary with polar angle indicating the presence of non-spherical aberration terms. A wavefront correction plate with rotationally anisotropic thickness is proposed to compensate for anisotropy in order to achieve good focusing by CRLs at beamlines operating at diffraction-limited storage rings.
Keywords: X-ray optics; wavefront correction; X-ray lenses; 3D printing; knife-edge imaging.
1. Introduction
Phase error correction in X-ray optics is a fast-evolving area of enabling technology
to generate pseudo perfect optics. The correction introduced by a suitable scheme
converts an aberrated optics to pseudo-perfect optics which otherwise prevents achieving
diffraction-limited focusing. A few schemes such as active bimorph mirrors (Mimura
et al., 2010), refractive correctors (Sawhney et al., 2016
; Seiboth et al., 2017
), invariable-multilayer deposition (Matsuyama et al., 2018
), diffractive wavefront correction (Probst et al., 2020
) and layer stress controlling method (Cheng & Zhang (2019
) have been demonstrated as tools for phase error corrections of different X-ray optical
elements. Refraction-based correctors are thin, easy to insert into the beam path,
do not change the optical axis and are straightforward to align. X-ray LIGA fabricated
SU-8 wavefront correctors were used in the wavefront error compensation of X-ray mirrors
(Laundy et al., 2017
) and X-ray LIGA fabricated lenses (Sawhney et al., 2019
) in one-dimensional (1D) focusing geometry. A silica refractive phase plate manufactured
by the laser ablation process was instrumental in reducing the phase error of two-dimensional
(2D) focusing Be CRLs (Seiboth et al., 2017
). In each case, the wavefront error is reduced due to the use of suitable correctors,
and our group recently demonstrated r.m.s. wavefront error compensation down to the
order of λ/100 (Laundy et al., 2019
).
Nano- and micro-fabrications have played a pivotal role in the development of novel
micro X-ray optical elements which led to a significant advance in achieving nano-
and micrometre-size focused X-ray beams (Li et al., 2020; Dhamgaye et al., 2014
; Lyubomirskiy, Boye et al., 2019
; Yan et al., 2014
). Highly sensitive X-ray optical measurements, especially sensitive wavefront error
measurements, were possible due to microfabrication of 1D and 2D X-ray gratings (Liu
et al., 2018
; Weitkamp et al., 2005
; Rutishauser et al., 2012
). Lithography techniques including Si etching, X-ray lithography and laser ablation
were used in the development of nano-focusing lenses and wavefront corrector plates.
Additive manufacturing or three-dimensional (3D) printing technology is developing
rapidly and is revolutionizing many key areas of industries and research. The 3D printer
is based on the two-photon polymerization process (Photonic Professional GT2 datasheet,
https://www.nanoscribe.com/fileadmin/Nanoscribe/Solutions/Photonic_Professional_GT2/DataSheet_PPGT2.pdf, NanoScribe GmbH), which is capable of patterning arbitrary 3D shapes with micrometre
or nanometre resolution. This printer was employed in many state-of-the-art device
developments (Dietrich et al., 2018
) and was recently used in X-ray optics developments (Sanli et al., 2018
; Petrov et al., 2017
; Lyobomirskiy, Koch et al., 2017
). The same 3D printer is used for the development of corrector plates.
X-ray optical elements, X-ray mirrors based on reflection, X-ray lenses based on refraction
and X-ray zone-plate/multilayer Laue lenses based on diffraction principles are used
for micro- or nano-focusing of the X-rays (Ice et al., 2011). The (n) in the X-ray region for X-rays with energy E is
where 1 − δ is a real term, δ is the index of refractive decrement (10−5 to 10−7) and β is an imaginary term that causes absorption (10−7 to 10−9). The real part of n is slightly less than unity in the X-ray region for all materials, thus the shape
of the X-ray lenses is concave, in contrast to convex used in the visible region.
Due to weak refraction power, multiple X-ray lenses are used in series by compounding
the refraction power of the lenses to achieve a reasonable focal length. Such X-ray
lens assemblies are known as compound refractive lenses (CRLs) (Snigirev et al., 1996). Parabolic-shaped 2D focusing X-ray lenses are fabricated in Be or Al by the mechanical
punching method, and 1D focusing lenses in Si, diamond or polymer are manufactured
by lithography techniques. Lens fabrication errors result in a deviation of the X-ray
pathlength in the lens from the ideal parabolic function. This causes a perturbation
of the X-ray wavefront which, when propagated to the focal plane, degrades the focus.
With respect to Be CRL fabrication, factors such as mechanical punching (two punching
tools with angular or spatial error), density variation, or variation in the chemical
composition of the material are responsible for the origin of wavefront errors. The
intensity distribution or wavefront errors of given optics are measured by a suitable
wavefront-sensing technique. Zernike polynomial fitting is a useful tool in diagnosing
visible optics wave aberrations over a circular or annular aperture. Zernike polynomials
expansion is used over the wavefront error map in quantifying optics aberrations present
in X-ray optics (Celestre et al., 2020
; Seiboth et al., 2016
; Zhou et al., 2018
). An imperfect optics produces blurred images of a source, and the performance improvement
of optics by aberration compensation schemes can be expressed in terms of reduction
in the coefficient of classical primary (Seidel) optics aberrations closely represented
by low-order Zernike polynomials.
Recently (Seaberg et al., 2019), a 3D printed phase plate in IP-S resist was used to correct wavefront errors of
20 Be lenses, and wavefront analysis was carried out using three wavefront reconstruction
techniques for X-ray free-electron laser (XFEL) sources. This paper describes the
use of the knife-edge imaging-based wavefront-sensing technique to determine wavefront
errors from two different stacks of Be lenses. This wavefront sensing technique is
described in our previous work (Laundy et al., 2019
). The optical characterization of a rotationally invariant profiled polymer corrector
plate manufactured by 3D printing was carried out at the Diamond Test beamline. After
correction with the phase plate, the r.m.s. wavefront error of 98 Be lenses showed
a reduction by a factor of six. The knife-edge imaging-based wavefront-sensing technique
was originally developed to measure 1D wavefront error profiles but in the present
studies it was adapted to measure the full rotational variant wavefront error profiles
of the Be CRL. The previously reported study (Seiboth et al., 2017
) used a rotationally invariant corrector and showed a reduction of spherical deformation
of the Be lenses. The present study reports the existence of a range of lower- and
higher-order optics aberrations in Be CRLs including spherical aberration, astigmatism
and coma. We highlight, particularly, that it is impossible to correct all optics
aberrations of X-ray lenses with a rotationally invariant corrector when rotationally
variant wavefront errors are present. A case study of the effect of rotationally invariant
corrector plates versus rotationally variant corrector plates on the corrected wavefront
error is described and evaluated in terms of Zernike polynomials. The r.m.s. wavefront
error is used to characterize the aberration level from the optics. X-ray lenses differ
from visible-light lenses in having weak refraction and strong absorption. This limits
the numerical aperture to of order ∼10−3. We present a modified form of the r.m.s. wavefront error with weighting due to transmitted
intensity.
2. Optical characterization setup
Diamond's Test beamline B16 was used for at-wavelength characterization of the X-ray
lenses and 3D printed corrector plate (Sawhney et al., 2010). A typical experimental setup used for the wavefront error measurement of CRLs is
shown in Fig. 1
. The monochromatic beam from a Si(111) double-crystal monochromator was focused by
X-ray lenses and observed on an X-ray detector placed at a distance of ∼1–2 m downstream
of the lens's focus. A corrector plate mounted on an alignment stage was positioned
in front of the Be CRLs. The Be CRLs (fabricated by RX Optics), knife-edge (fabricated
by X-ray LIGA at ANKA synchrotron) and X-ray detectors (Mini-FDS from Photonics Science
and PIPS diode) were mounted on stable rigid platforms. A CRL consisting of N = 98 individual Be bi-concave parabolic-shaped lenses, 200 µm radius of curvature
at the apex and theoretical focal length 673 mm (image distance q = 696 mm) at 15 keV was installed. We will refer to this CRL set as CRL1. A 2D pixel
area detector Mini-FDS of pixel size 6.45 µm was used to record the images as a function
of knife-edge position. X-ray transmission of the corrector plate was measured using
a PIPS diode. A second set of X-ray lenses with N = 24 (referred to here as CRL2) was characterized in the same setup with revised
positions of knife-edge and detectors from the centre of the CRL.
![]() |
Figure 1 (a) Schematic experimental setup at B16 Test beamline, Diamond Light Source. (b) The geometry used for rotationally variant wavefront error measurements; a knife-edge was rotated at an angle (e.g. ϕ = 45°) for the measurement of the corresponding wavefront errors in the lens area highlighted by the pink stripe. |
The wavefront error measurement involved recording the (b). The 1D wavefront error along the diameter of the lenses, i.e. the central line of the shaded area inclined at an angle, was measured by orientating
the knife-edge at the same angle and collecting intensity data of the recorded image
from a narrow strip of pixels tilted at the same angle. The wavefront error measured
along the diameter was resolved into two radial functions (0 to r) separated by 180° around the polar axis, i.e. two radial wavefront profiles at 45° and 225° for a 45° knife-edge orientation. The
centre of the lenses is located on the detector as the position of maximum intensity
transmission. A wavefront error that is constant as a function of radial distance
of the entire polar angles is called a rotationally invariant wavefront error. Similarly,
a wavefront error that varies as a function of entire polar angle as well as radial
distance is called a rotationally variant wavefront error. The optimum performances
of the 3D printed corrector plate were analysed by comparing the r.m.s. wavefront
errors of Be lenses with and without a corrector plate and comparing focused beam
sizes in two orthogonal planes near to the optics focal plane. We define the r.m.s.
wavefront error for the lenses over its aperture of radius R0 weighting with the transmitted intensity as
where the wavefront error w(r,ϕ) is weighted by the I(r) which, for uniform incident intensity over the lens aperture I0, is given by the [μ(E)],
where RL is the radius of curvature of the lens, E is the X-ray energy, N is the number of the lenses, and the wavefront error w(r,ϕ) is defined on an aperture 0 ≤ r ≤ R0 and 0 ≤ ϕ ≤ 2π.
The wavefront error over a circular aperture can be expressed as a series of Zernike
polynomials as functions of the normalized radial position r/R0 and radial angle 0–2π. These are a complete set of basis functions that are orthogonal over a circle of
unit radius and are commonly used to represent optical aberrations (Born & Wolf, 1999). The Zernike polynomials in the Noll notation which uses a single index j are defined as (Noll, 1976
)
where m is the azimuthal frequency, n is the radial degree, ρ = r/R0 and
The index j is the mode ordering number which is expressed in terms of n and m. The Zernike polynomial modes (Zj) expansion of the Be lens arbitrary wavefront error is expressed as =
where Zj is the Zernike coefficient for each Zj obtained from
and
The Zernike polynomial Python library provided by Fan (2019) is used for fitting lens wavefront errors and determining Zernike coefficients.
3. Corrector plate design
Measurement of the figure error distribution in the Be CRLs is required for the design
of a corrector plate. An ideal coherent wavefront from the source at 47 m upstream
was considered at the entrance of the Be CRLs. For ideal lenses, an emerging wavefront
at the exit of the lenses will be a converging spherical wavefront radius centred
on the focus. In reality, the emerging wavefront from the Be CRLs is distorted by
variation of lens thickness from the ideal parabolic profile caused by imperfect manufacturing.
Other factors such as impurity in the lens material or non-uniform pressed lens material
during manufacturing leading to density variations contribute to the origins of wavefront
errors of the optic. A knife-edge imaging technique is used for the first time for
the investigation of figure error distribution in Be CRLs. This technique reproduces
measurements for the particular optics, and wavefront errors recorded for Be lenses
are found on a similar order as measured by the other techniques, e.g. ptychography or speckle tracking (Seaberg et al., 2019). The 1D measured wavefront errors along vertical and horizontal lines are shown
in Fig. 2
for four different polar angles as a function of radial position. The lenses are
randomly oriented in the casing and show different wavefront error functions at different
polar angles. An invariant wavefront profile around the polar axis is evident for
the polar angles 90°and 270° (green solid and dashed lines in Fig. 2
) which was measured whilst the knife-edge was stepped along the horizontal line.
However, we considered an average error profile calculated from the error profiles
measured at different polar angles (solid blue line) and the same error is converted
into the design of a rotationally invariant corrector. An optical path length difference
(Δw) is introduced by a material of thickness (t) with phase error (θ),
where t = and δ is the refractive decrement of the X-ray (n) given in equation (1)
. The ratio δ(E)/β(E) can be used as a selection criterion for choosing a corrector plate material with
higher ratio of refraction power to X-ray absorption. Thus, low-atomic-number materials
are preferred over higher-atomic-number materials. Materials such as Be, Al, Si, diamond
and polymers composed of carbon, hydrogen and oxygen are commonly used for micro X-ray
optical elements. A polymer-based corrector plate is used in the present study and
its thickness required for compensation wavefront error is calculated using equation
(7)
. A typical 3D printable polymer IP-S of thickness difference Δt = 10 µm will produce a phase advance 2πδΔt/λ and will introduce an optical path difference of 11.74 pm [molecular formula C14H18O7, density = 1.2 g cm−3 (Lyubomirskiy, Koch et al., 2019
)]. An estimated thickness profile of the IP-S corrector in a 3D symmetry is shown
in Fig. 2
(b). Many 3D printers based on fusion deposition modelling or stereolithography produce
structures with feature size >>1 µm with a high degree of porosity in the fabricated
structure. A nanoscribe 3D printer is an ideal tool for 3D printing of the corrector
plate (Nanoscribe GmbH, Germany). It is capable of printing arbitrary features with
sub-micrometre precision in three dimensions. The surface finish of the printed structure
is ∼20 nm which is good for the normal-incidence optics used in the X-ray region (Photonic
Professional GT2 datasheet, https://www.nanoscribe.com/fileadmin/Nanoscribe/Solutions/Photonic_Professional_GT2/DataSheet_PPGT2.pdf).
![]() |
Figure 2 (a) Wavefront error distribution of 98 Be lenses in its 2D circular aperture at various polar angles. The solid blue line shows the average wavefront error over polar angles measured and was used for the design of a rotationally invariant corrector plate. (b) Schematic 3D design of the corrector plate. |
The design of the corrector plate was prepared in AutoCAD and converted into a 3D CAD-step file. A corrector plate was fabricated using Nanoscribe Photonic Professional GT2 (Nanoscribe GmbH, Germany) at Lancaster University, UK. The dip in lithography mode was used to print the 3D design in IP-S (Nanoscribe GmbH). The laser source used for printing was a femtosecond Ti-sapphire type (800 nm, 80 MHz, 50 fs). IP-S was drop-cast onto ITO (75–100 Ω2) coated glass N1.5 thickness coverslips (Diamond Coatings Ltd, Halesowen, UK). The resist was exposed from bottom to top using a femtosecond laser pulse focused in voxel by 25× objective with laser power 55% and writing speed 200000 µm s−1. Patterned IP-S resist was developed in PGMEA for 20 min, rinsed in IPA for 5 min and dried using N2 enriched air.
4. Results and discussions
4.1. Rotationally invariant wavefront errors measurement and its compensation
The effectiveness of the corrector plate in the wavefront error compensation depends
on various factors such as repeatability of wavefront measurements between successive
beam times, the stability of the optics/beam, design of the corrector plate, fabrication
errors in the corrector plate and alignments of the corrector plate to the CRL optic
axis. A rotationally invariant 3D printed corrector was placed upstream to the Be
CRL as shown in Fig. 1 for the figure error corrections of the Be CRL. The wavefront errors of Be CRL1 were
measured and the repeatability in the measurements was confirmed by comparing the
measurement with that made during the design of the corrector plate. Good alignment
of the centre of the corrector plate relative to the lens optical axis in a beam path
is critical in achieving optimum compensation results. With the corrector plate position
in the nearly plane wavefront before the focusing lenses, the correction is insensitive
to the correctors' longitudinal position. The lateral position of the phase plate
is more important, with good alignment to the axis of the lens being required. To
achieve this, the phase plate was stepped laterally within the lens aperture with
coarser 5 µm and finer 1 µm step size and the corresponding r.m.s. wavefront error
was determined using equation (2)
. The best lateral positions for the corrector plate are achieved by minimizing the
r.m.s. wavefront error in the respective planes.
An average of the CRL1 wavefront errors measured at four different polar angles before
and after the corrections is shown in Fig. 3. The r.m.s. wavefront error [equation (2)
] is found to be 14.4 pm before the correction and 2.4 pm after the correction which
is an improvement by a factor of six. The expected performance of a designed corrector
plate is shown as `after correction (calculated)' in Fig. 3
which is obtained by subtracting the wavefront error values used for the design of
the corrector plate (dashed magenta) from the corresponding error values measured
for CRL1 before correction (blue). The r.m.s. wavefront error difference between the
designed corrector (discussed in Section 2
) and the fabricated corrector is <1 pm. This difference is due to various contributions
such as infidelity in corrector fabrication, alignment/stability of optics and repeatability
in the wavefront measurements. X-ray absorption by the corrector was calculated by
measuring the PIPS diode photocurrent for direct beam and placing the corrector plate
in the beam path. The transmission of the corrector plate was found to be ∼99%. A
clear improvement in the focus profiles in the vertical and horizontal direction was
observed after the introduction of the corrector [Figs. 4
(a) and 4
(b)]. The focus profiles, before and after corrections, are measured at the same focal
distance from the centre of the CRL. The corrector plate has improved the vertical
(horizontal) focus size to 0.9 µm (2.5 µm) from 2.3 µm (3.7 µm) due to the aberrated
wavefront. The focus size of CRLs at a bending-magnet source is limited by the size
of the de-magnified source.
![]() |
Figure 3 Rotationally invariant wavefront errors of the Be CRLs, before correction, after correction and wavefront error profile by rotationally invariant corrector plate. The r.m.s. wavefront error calculated in each case is given in square parentheses. |
![]() |
Figure 4 Improvement in imaging the B16 bending-magnet source in the (a) vertical and (b) horizontal direction after insertion of the corrector plate. Solid lines show a Gaussian fit for the corresponding focus profile measurements. |
A type of wavefront aberration exists in CRL1 before and after the corrections were
quantified using Zernike polynomials expansion up to order n = 16. Fig. 5 shows the amplitude of the first 36 Zernike coefficients and coefficients corresponding
to higher-order spherical aberrations only (Z37, Z56, Z79, Z106 and Z137) as the values of the remaining coefficients of the higher orders are either small
or zero. Zernike polynomial coefficients Z1 to Z4 are not aberrations but they describe the surface positioning. Z1 is constant over the whole aberration map and therefore not considered. The misalignment
of optics is expressed in the system tilts Z2 and Z3 along two orthogonal planes and term Z4 defines defocusing. The major optics aberrations observed in the Be CRLs were due
to primary (Z11), secondary (Z22), tertiary (Z37) and higher-order spherical aberrations. These spherical deformations were well
corrected after the introduction of the corrector plate. The defocus term (Z4) observed was caused by the displacement of the knife-edge from the focal plane
in the direction along the optical axis. This study does not show a contribution from
non-spherical aberration terms. The r.m.s. wavefront error is given as the sum of
squares of all Zernike coefficients. The r.m.s. calculated by considering all Zernike
coefficient values except (Z1–Z4) is 14.2 pm before correction and 2.7 pm after correction. These values match well
with the ones calculated using equation (2)
.
![]() |
Figure 5 Zernike polynomial fitting over measured and averaged wavefront errors of CRL1 before and after correction (measured and calculated corrector plate contribution). |
4.2. Rotationally variant wavefront errors measurement and its compensation
We extended our investigation to another set of lenses: CRL2 (N = 24). We investigated the polar-angle-resolved wavefront error distributions by
making wavefront measurements with the knife-edge rotated in angles about the optical
axis to obtain the radial wavefront error over the polar angle from 0° to 360°. The
intensity recorded in a 2D pixel detector was processed only for those pixels that
lie along a line inclined at a rotated angle. Unfortunately, the knife-edge scan data
is not complete for CRL1 as the measurement script failed twice during the experiments.
An average radial wavefront error calculated over a complete radial profile was used
for missed measurements at the polar angles (135–165° and 315–345°). Figs. 6(a) and 6(b) show polar plots of the wavefront errors in both CRLs before correction. The wavefront
errors of both CRLs are close to being invariant but show anisotropic wavefront error
distributions in the polar angles. The distributions are not radially concentric but
approximately oval, rotated at 45° and 90° for CRL1 and CRL2, respectively. An analytical
approach was considered to evaluate the performance of the rotationally invariant
corrector plate in compensating for the rotationally variant wavefront errors of CRL1
and CRL2. The remaining wavefront errors after correction by the rotationally invariant
corrector plates are shown in Figs. 6
(c) and 6(d).
![]() |
Figure 6 Polar plots showing rotationally variant wavefront errors of Be CRLs (a, b) and after correction using rotationally invariant corrector plates (c, d) for CRL1 and CRL2, respectively. |
The uncorrected wavefront errors of both CRLs were found in a similar range. We noticed no per-lens wavefront error accumulation – otherwise peak-to-peak wavefront errors of CRL1 would be four times higher than for CRL2 over the whole lens aperture. This observation is true near the optical axis of the lenses where maximum transmission of the X-rays is observed. Any rotation of the individual lens in the lens casing may be averaging figure errors and such averaging is apparent more in CRL1 compared with CRL2.
The wavefront error surfaces shown in Figs. 6(a)–6(d) were fitted with Zernike polynomials, and corresponding amplitudes of Zernike coefficients
are shown in the bar chart in Fig. 7
. To avoid areas of non-measurements in the fitting and obtain a good fit, a radial
distance (R0) of ±186 µm for CRL1 and ±305 µm for CRL2 from the centre of the wavefront error
map was chosen. The strength of various optics aberration expressed by Zernike polynomials
expansion before and after corrections shows the existence of lower and higher orders
of spherical and non-spherical optics aberrations. As discussed in the previous section,
here too spherical aberrations of both CRLs are compensated well by the rotationally
invariant corrector plate. However, non-spherical aberration terms (including astigmatism,
coma, etc.) and higher-frequency terms (trefoil, tetrafoil, pentafoil, hexafoil, etc.) remained uncorrected. Astigmatism in CRL2 contributes significantly to the remaining
optics aberration which cannot be ignored for obtaining diffraction-limited focusing.
The primary optics aberration tilt, defocus, astigmatism, coma, and spherical aberration
are expressed in terms of Zernike coefficients (Z2, Z3), Z4, (Z5, Z6), (Z7, Z8) and Z11, respectively.
![]() |
Figure 7 Zernike polynomial fitting over measured rotationally variant wavefront errors of CRL1 and CRL2 before and after correction (calculated rotationally invariant corrector plate contribution). |
We propose two possibilities (case 1 and case 2) for correction of X-ray optics aberrations in CRLs using customized corrector plates. In the first case, a corrector plate is fabricated with a thickness profile in two dimensions that fully corrects the wavefront over the full aperture of the lens. In the second case the spherical terms are corrected using a radially invariant corrector and an additional in-line corrector plate used to correct selected radially variant higher-order terms in the Zernike expansion, such as astigmatic terms.
4.2.1. Case 1
This has the advantage that complete correction can be achieved with a single corrective element; however, alignment becomes more difficult, as in addition to transverse alignment the corrective optic must also be aligned in rotation angle about the optical axis. It is also necessary to measure the full 2D wavefront error in order to design the profile of the corrector.
We have worked out the designs of such rotationally variant corrector plates exclusively
for CRL1 and CRL2. The designs are shown in Figs. 8(a) and 8(b) and they can be fabricated by 3D printing. Such corrector plates are planned to
be manufactured in the near future and characterized using CRL1 and CRL2 at the Diamond
Test beamline. The proposed rotationally variant corrector plate can be extended for
the wavefront corrections of CRLs made from Al or polymer materials. An exclusive
3D corrector plate is feasible to build on the same chip in line with the nano-focusing
lenses fabricated by either LIGA or semiconductor manufacturing techniques. The 3D
correctors made in IP-S polymer are useful at a bending-magnet source, but this polymer
degrades quickly in the higher-intensity beams of undulator or XFEL sources. However,
searching for a robust material for the corrector plate is necessary to deploy rotationally
invariant/variant corrector plates with X-ray optics at beamlines operational on diffraction-limited
storage rings or XFELs.
![]() |
Figure 8 Estimated design of a rotationally variant corrector plate for (a) CRL1 and (b) CRL2. |
4.2.2. Case 2
In CRL2, the low-order astigmatism (Zernike polynomials Z5 and Z6) contributes almost 50% of the total remaining aberrations after the correction introduced by the radially invariant corrector plate. A customized second corrector plate dedicated to the compensation of low-order astigmatism can be designed in the following way. The wavefront error due to astigmatism is
and in the Cartesian form
where Z5 and Z6 are Zernike coefficients for Zernike mode Z5 and Z6 and their values are extracted from fitted data (Fig. 7), and a and b are constants.
The above wavefront definition creates a parabolic surface in the xy plane and it is possible to manufacture using a 3D printer. Fabrication of a sequence
of correctors would allow a degree of adaptability to be incorporated into the correction.
Table 1 summarizes the coefficients of the lower-order Zernike polynomials that closely represent
classical aberrations, for Be CRLs wavefront errors before and after corrections [`Corrector1',
rotationally invariant corrector plate; `Corrector2', rotationally variant corrector
plate, as defined in equation (9)
]. The r.m.s. wavefront error of the optics is reduced from 24.0 pm to 13.3 pm with
the first-order correction plate and finally to 4.69 pm (∼0.06λ) with the second-order correction plate. For primary aberrations of CRL2 excluding
the piston, tilt and defocus terms, the obtained r.m.s. value is ∼1 pm.
|
5. Conclusions
The knife-edge imaging wavefront-sensing technique was successfully used in X-ray lenses wavefront error measurements and the optical characterization of a 3D printed corrector. The use of a rotationally invariant 3D printed wavefront corrector plate in wavefront errors compensation of 98 Be X-ray lenses was demonstrated. The r.m.s. wavefront error of rotationally invariant wavefront aberrations in Be CRLs was reduced by 84% after the introduction of a rotationally invariant 3D printed corrector. Zernike polynomials analytical fitting is useful in the quantification of optics aberrations before and after correction wavefront errors. All orders of spherical aberrations are found corrected after the insertion of a rotationally invariant corrector plate but it is apparent that significant non-spherical aberrations still remain. Thus, a rotationally invariant corrector plate is unable to completely compensate optics aberrations CRLs. The knife-edge imaging technique was adapted to measure the full 2D wavefront errors of two X-ray lenses sets CRL1 and CRL2. The Zernike polynomial fitting of measured wavefront error maps of CRL1 and CRL2 showed the existence of lower- and/or higher-order rotationally invariant and variant optics aberrations. We have therefore specified wavefront corrector plates which could approach complete compensation of the wavefront errors. The role of the present 3D printer technology is important in achieving the precision manufacturing of rotationally variant corrector plates. This is a possible way to tackle optics aberrations in X-ray optics and achieving r.m.s. wavefront error compensation below 0.07λ. The present framework of wavefront measurement and corrections is useful in X-ray optics being used at the third- and fourth-generation synchrotron facilities and XFELs.
Acknowledgements
This work was carried out with the support of Diamond Light Source. The authors thank Oliver Fox and Andrew Malandain for the help and support during the measurements on the Diamond Test beamline. We are grateful to Peter Docker for introducing the Lancaster University Group for 3D printing and thankful to Arndt Last from IMT/KIT for providing the Au knife-edge used in the knife-edge imaging technique.
Funding information
The following funding is acknowledged: European Union's Horizon 2020 research and innovation programme under the Marie Sklodowska-Curie Actions awarded to the Science and Technology Facilities Council, UK (grant No. 665593).
References
Born, M. & Wolf, E. (1999). Principles of Optics. Cambridge University Press. Google Scholar
Celestre, R., Berujon, S., Roth, T., Sanchez del Rio, M. & Barrett, R. (2020). J. Synchrotron Rad. 27, 305–318. Web of Science CrossRef IUCr Journals Google Scholar
Cheng, X. & Zhang, L. (2019). J. Synchrotron Rad. 26, 234–237. CrossRef IUCr Journals Google Scholar
Dhamgaye, V. P., Tiwari, M. K., Garg, C. K., Tiwari, P., Sawhney, K. J. S. & Lodha,
G. S. (2014). Microsyst Technol. 20, 2055–2060. CrossRef Google Scholar
Dietrich, P. I., Blaicher, M., Reuter, I., Billah, M., Hoose, T., Hofmann, A., Caer,
C., Dangel, R., Offrein, B., Troppenz, U., Moehrle, M., Freude, W. & Koos, C. (2018).
Nat. Photon. 12, 241–247. CrossRef CAS Google Scholar
Fan, X. (2019). Python Optics Module, MIT Licence, https://github.com/Sterncat/opticspy. Google Scholar
Ice, G. E., Budai, J. D. & Pang, J. W. (2011). Science, 334, 1234–1239. Web of Science CrossRef CAS PubMed Google Scholar
Laundy, D., Dhamgaye, V., Moxham, T. & Sawhney, K. (2019). Optica, 6, 1484–1490. Web of Science CrossRef Google Scholar
Laundy, D., Sawhney, K., Dhamgaye, V. & Pape, I. (2017). Proc. SPIE, 10386, 103860B. Google Scholar
Li, K., Ali, S., Wojcik, M., De Andrade, V., Huang, X., Yan, H., Chu, Y. S., Nazaretski,
E., Pattammattel, A. & Jacobsen, C. (2020). Optica, 7, 410–416. CrossRef CAS Google Scholar
Liu, Y., Seaberg, M., Zhu, D., Krzywinski, J., Seiboth, F., Hardin, C., Cocco, D.,
Aquila, A., Nagler, B., Lee, H., Boutet, S., Feng, Y., Ding, Y., Marcus, G. & Sakdinawat,
A. (2018). Optica, 5, 967–975. CrossRef CAS Google Scholar
Lyubomirskiy, M., Boye, P., Feldkamp, J. M., Patommel, J., Schoeder, S., Schropp,
A., Burghammer, M., Wild, C. & Schroer, C. G. (2019). J. Synchrotron Rad. 26, 1554–1557. CrossRef CAS IUCr Journals Google Scholar
Lyubomirskiy, M., Koch, F., Abrashitova, K., Bessonov, V., Kokareva, N., Petrov, A.,
Seiboth, F., Wittwer, F., Kahnt, M., Seyrich, M., Fedyanin, A., David, C. & Schroer,
C. G. (2019). Opt. Express, 27, 8639–8650. CrossRef CAS PubMed Google Scholar
Matsuyama, S., Inoue, T., Yamada, J., Kim, J., Yumoto, H., Inubushi, Y., Osaka, T.,
Inoue, I., Koyama, T., Tono, K., Ohashi, H., Yabashi, M., Ishikawa, T. & Yamauchi,
K. (2018). Sci. Rep. 8, 17440. Web of Science CrossRef PubMed Google Scholar
Mimura, H., Handa, S., Kimura, T., Yumoto, H., Yamakawa, D., Yokoyama, H., Matsuyama,
S., Inagaki, K., Yamamura, K., Sano, Y., Tamasaku, K., Nishino, Y., Yabashi, M., Ishikawa,
T. & Yamauchi, K. (2010). Nat. Phys. 6, 122–125. Web of Science CrossRef CAS Google Scholar
Noll, R. J. (1976). J. Opt. Soc. Am. 66, 207–211. CrossRef Google Scholar
Petrov, A., Bessonov, V., Abrashitova, K., Kokareva, N., Safronov, K., Barannikov,
A., Ershov, P., Klimova, N. B., Lyatun, I. I., Yunkin, V. A., Polikarpov, M., Snigireva,
I., Fedyanin, A. A. & Snigirev, A. (2017). Opt. Express, 25, 14173–14181. CrossRef CAS PubMed Google Scholar
Probst, J., Braig, C., Langlotz, E., Rahneberg, I., Kühnel, M., Zeschke, T., Siewert,
F., Krist, T. & Erko, A. (2020). Appl. Opt. 59, 2580–2590. CrossRef PubMed Google Scholar
Rutishauser, S., Samoylova, L., Krzywinski, J., Bunk, O., Grünert, J., Sinn, H., Cammarata,
M., Fritz, D. M. & David, C. (2012). Nat. Commun. 3, 947. Web of Science CrossRef PubMed Google Scholar
Sanli, U. T., Ceylan, H., Bykova, I., Weigand, M., Sitti, M., Schütz, G. & Keskinbora,
K. (2018). Adv. Mater. 30, 1802503. Web of Science CrossRef Google Scholar
Sawhney, K., Dolbnya, I., Tiwari, M., Alianelli, L., Scott, S., Preece, G., Pedersen,
U., Walton, R., Garrett, R., Gentle, I., Nugent, K. & Wilkins, S. (2010). AIP Conf. Proc. 1234, 387–390. CrossRef CAS Google Scholar
Sawhney, K., Laundy, D., Dhamgaye, V. & Pape, I. (2016). Appl. Phys. Lett. 109, 051904. Web of Science CrossRef Google Scholar
Sawhney, K., Laundy, D., Moxham, T., Fox, O. & Dhamgaye, V. (2019). AIP Conf. Proc. 2054, 060003. Google Scholar
Seaberg, M., Cojocaru, R., Berujon, S., Ziegler, E., Jaggi, A., Krempasky, J., Seiboth,
F., Aquila, A., Liu, Y., Sakdinawat, A., Lee, H. J., Flechsig, U., Patthey, L., Koch,
F., Seniutinas, G., David, C., Zhu, D., Mikeš, L., Makita, M., Koyama, T., Mancuso,
A. P., Chapman, H. N. & Vagovič, P. (2019). J. Synchrotron Rad. 26, 1115–1126. Web of Science CrossRef CAS IUCr Journals Google Scholar
Seiboth, F., Kahnt, M., Scholz, M., Seyrich, M., Wittwer, F., Garrevoet, J., Falkenberg,
G., Schropp, A. & Schroer, C. G. (2016). Proc. SPIE, 9963, 99630P. Google Scholar
Seiboth, F., Schropp, A., Scholz, M., Wittwer, F., Rödel, C., Wünsche, M., Ullsperger,
T., Nolte, S., Rahomäki, J., Parfeniukas, K., Giakoumidis, S., Vogt, U., Wagner, U.,
Rau, C., Boesenberg, U., Garrevoet, J., Falkenberg, G., Galtier, E., Ja Lee, H., Nagler,
B. & Schroer, C. (2017). Nat. Commun. 8, 14623. CrossRef PubMed Google Scholar
Snigirev, A., Kohn, V. G., Snigireva, I. & Lengeler, B. (1996). Nature, 384, 49–51. CrossRef CAS Web of Science Google Scholar
Weitkamp, T., Nöhammer, B., Diaz, A., David, C. & Ziegler, E. (2005). Appl. Phys. Lett. 86, 054101. Web of Science CrossRef Google Scholar
Yan, H., Conley, R., Bouet, N. & Chu, Y. S. (2014). J. Phys. D Appl. Phys. 47, 263001. Web of Science CrossRef Google Scholar
Zhou, T., Wang, H., Fox, O. & Sawhney, K. (2018). Opt. Express, 26, 26961–26970. CrossRef CAS PubMed Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.