research papers
Challenges and solutions for the analysis of in situ, in crystallo micro-spectrophotometric data
aSwiss Light Source, Paul Scherrer Institute, CH-5232 Villigen PSI, Switzerland,bSchool of Biological Sciences, University of Essex, Wivenhoe Park, Colchester CO4
3SQ, England,cEuropean Molecular Biology Laboratory Hamburg, c/o DESY, Notkestrasse 85, D-22603
Hamburg, Germany, and dPhoton Sciences, Brookhaven National Laboratory, Mail Stop 745, Upton, NY 11973, USA
*Correspondence e-mail: [email protected]
Combining macromolecular crystallography with in crystallo micro-spectrophotometry yields valuable complementary information on the sample, including the redox states of metal cofactors, the identification of bound ligands and the onset and strength of undesired also known as radiation damage. However, the analysis and processing of the resulting data differs significantly from the approaches used for solution spectrophotometric data. The varying size and shape of the sample, together with the suboptimal sample environment, the lack of proper reference signals and the general influence of the X-ray beam on the sample have to be considered and carefully corrected for. In the present article, how to characterize and treat these sample-dependent artefacts in a reproducible manner is discussed and the SLS-APE in situ, in crystallo data-analysis toolbox is demonstrated.
Keywords: micro-spectrophotometer; UV–visible absorption; Raman spectroscopy; single-crystal spectroscopy; software toolbox; data analysis; background correction.
1. Introduction
Complementary biophysical techniques are a powerful tool to complete and enhance structural
information obtained by macromolecular crystallography (MX). The structural information
provided by electron-density maps obtained from diffraction experiments often does
not provide sufficient insight to answer the point of interest conclusively. This
is especially true in the fields of enzyme kinetics and protein function, ligand-binding
modes or redox states of metal centres. In situ micro-spectrophotometry is a highly effective way to obtain such complementary information
about the chemical identity of cofactors, the electronic state of metal centers or
about specific radiation-induced chemistry. Since its first introduction in the 1970s
(Rossi & Bernhard, 1970) it has come a long way, and today is implemented in several experimental endstations
of macromolecular crystallography beamlines at storage rings around the globe, allowing
a larger community of structural biologists to benefit from the opportunities provided
by this complementary method. Examples include instruments at the ESRF (Carpentier
et al., 2007
; Royant et al., 2007
; Davies et al., 2009
), APS (Pearson et al., 2007
), NSLS (Stoner-Ma et al., 2011
; Orville et al., 2011
), DLS (Allan et al., 2013
), SPring-8 (Sakai et al., 2002
; Shimizu et al., 2013
) and SLS (Owen et al., 2009
; Pompidor et al., 2013
). Spectroscopies successfully applied to protein crystals in situ include UV–visible absorption, fluorescence, resonance Raman and nonresonance Raman
as well as X-ray absorption spectroscopy. A particularly useful application of complementary
techniques is the assessment of X-ray-induced photophysics in the sample, also known
as radiation damage. The combination of in situ spectroscopies with X-ray-induced photophysics allows the characterization of transient
states accessible by (Schlichting et al., 2000
; Hough et al., 2008
; Adam et al., 2009
; Hersleth & Andersson, 2011
), monitoring of the oxidation states of metal centres (Beitlich et al., 2007
; Ellis et al., 2008
; He et al., 2012
; Merlino et al., 2013
), quantification of radiation damage (Meents et al., 2007
; Rajendran et al., 2011
; McGeehan et al., 2007
) and the general understanding of its mechanisms (Carpentier et al., 2010
; Sutton et al., 2013
).
While synchrotron users are often proficient in interpreting solution spectroscopic data, the particular artefacts observed in the spectra of samples in crystalline form can be easily misinterpreted. Spectroscopic measurement series typically have to cover multidimensional parameter spaces, and a rapid overview of the collected data is paramount in making maximum use of precious beamtime. The provision of a freely available spectroscopic data-analysis toolbox will therefore be a useful aid for users to properly process and gain the maximum biochemical information from their in crystallo spectroscopic data and overall crystallographic experiment.
1.1. UV–visible absorption spectroscopy
Absorption spectroscopy has been widely applied to fingerprint the states of chromophores
in protein crystals and to monitor changes in these during diffraction experiments.
This method is particularly suitable where a relatively low spectroscopic sensitivity
but a high time resolution is required. The main disadvantage of the method is the
fact that it not only requires the presence of a chromophore in the sample but also
that this chromophore is affected by the reaction under observation. A large fraction
(roughly 20%) of all of protein structures deposited in the Protein Data Bank (PDB)
contain a chromophore (Orville et al., 2011). Based on the set of most common chromophores defined by Orville and coworkers,
one can estimate this fraction to have remained more or less constant over the exponential
growth of the PDB in recent years (Fig. 1
and Supporting Information). In summary, approximately one fifth of all PDB structures
are suitable in principle for absorption spectroscopy. However, only a very small
proportion of those samples have been investigated spectroscopically. The potential
of the additional information to be gained here is the main motivation behind the
efforts of researchers and the PDB to include the possibility of submitting spectroscopic
data to the PDB alongside the structural model (Garman & Weik, 2011
; Orville et al., 2011
).
![]() |
Figure 1 (a) Relationship between the total number of X-ray structures deposited in the PDB (red) and structures containing chromophores (blue). The fraction of chromophore-containing structures (yellow) shows only a minimal decrease over this period. (b) Distribution of chromophores in X-ray structures deposited in the PDB containing common chromophores as of January 2014. |
1.2. Raman and resonance Raman spectroscopy
In contrast to absorption spectroscopy, Raman spectroscopy (RS) does not require the
presence of a chromophore. Instead, any bond vibration of proper symmetry can transfer
or receive energy to or from a scattered photon, theoretically allowing most bonds
in the molecule to be probed. While this is in general an advantage, it can become
a problem in the case of complex molecules such as proteins, since the wealth of information
makes the interpretation of the resulting spectra a challenging task. The maximal
theoretical number of vibrational modes for a nonlinear molecule can be computed by
the formula 3n − 6, where n denotes the number of atoms present. In a crystal one has to consider not only the
single molecule but the number of molecules in the owing to the non-isotropic orientation. In the case of the tetragonal trapezoidal
hen egg-white lysozyme (HEWL; PDB entry 1lyz ; Diamond, 1974), which contains eight molecules in the this results in 46 992 possible modes. Even though not all possible vibrational modes
are Raman active, it is near-impossible to perform de novo peak assignment without additional information, for example that gained from isotope-labelling
experiments.
However, it is possible to selectively excite bond vibrations related to a chromophore
by utilizing light of a wavelength corresponding to the absorption band of that chromophore
(Harrand & Lennuier, 1946). This so-called resonance Raman (RR) effect makes the measurement and interpretation
of protein spectra significantly easier, as bands arising from the chromophore will
be of greatly increased intensity (by several orders of magnitude) and far fewer in
number than bands measured in nonresonance experiments. However, owing to the higher
energy absorption upon photo-irradiation this technique may also have destructive
effects on the sample. Thus, great care has to be taken in the balance between photo-excitation
and photo-emission to avoid sample alteration or bleaching of the sample. The use
of both methodologies for protein analysis has been well described (Rippon et al., 1971
; Spiro & Strekas, 1974
; Carey, 1978
, 1999
; Palings et al., 1987
; Thomas, 1999
). The clear disadvantage of the RR technique compared with RS is the requirement
for a chromophore within the protein crystal.
2. Limitations of in crystallo spectroscopy
2.1. Protein crystals as samples and the nature of in crystallo spectra
In addition to the limitations of the techniques described above, e.g. Mitegen UV-Vis Mounts) can further improve the spectral quality. A good way to mount a crystal, in particular for Raman measurements, is to use a loop significantly smaller than the sample so that the crystal sticks out enough to allow unhindered measurements. However, this may not be appropriate for very thin plate-like crystals, which are optimal for absorption spectroscopy, since the surface strain might cause bending.
on solids is in general no easy task. For absorption techniques the samples must be thin and transparent enough to pass light. Depending on the of the chromophore in question, even a 10 µm thick plate-like crystal might be too thick. Often the crystal is surrounded by a liquid or amorphous solid phase consisting of the crystallization buffer and cryoprotectant. The chemical components of these can contribute to the spectra and have to be carefully subtracted. Owing to the different diffractive indices of the crystal and the liquid, beam displacement might occur, making it difficult to correctly align the sample and amplifying stray light artifacts. Samples should ideally be prepared with as little buffer as possible on a support (mesh or loop) allowing unhindered illumination. Newly available UV-transparent mounting loops (Owing to the high scatter density and the imperfect surface structure (e.g. crystal layer grating; see Fig. 2), in solids can be up to 106 times stronger than in solution samples (Julien, 1980
; Ohana et al., 1986
), thus significantly decreasing the signal-to-noise ratio (SNR). Since conventional
single-monochromator spectrographs are usually able to attenuate stray light with
an efficiency of only about 10−5 (Ohana et al., 1986
; Kim et al., 2010
), the use of an additional dichroic filter or a double monochromator is mandatory
to achieve a sufficient SNR. It is also critical to find a good alignment of spectroscopic
axes towards the sample. This is important to minimize beam displacements on the surfaces
in absorption measurements and scattering away from the collection objective in scattering
techniques. Usually this has to be achieved by rotating the sample in the instrument
until good spectra can be acquired.
![]() |
Figure 2 Stray light generation by crystalline samples (see text). (a) Grating effect of crystal layers. (b) Refraction owing to change in refractive indices and owing to imperfect crystal surface. |
Since not only the protein in the crystal but essentially all substances irradiated by the excitation laser can emit Raman photons, contaminants, buffers or ligands can all contribute to the measured RS or RR spectrum. In fact, the resulting background signal can be some orders of magnitude more intense than the weak Raman signal arising from the protein. It is thus desirable to eliminate as many of these factors as possible from the sample. Since this is hardly possible in in crystallo spectroscopy, at least the Raman signatures of the individual components should be known or measured independently and subtracted from the measured spectrum. Fortunately, water is a poor Raman scatterer owing to the lack of polarizable bond vibrations, so that, in contrast to infrared spectroscopy, aqueous solutions are usually a better choice compared with organic solvents. Both RS and RR spectroscopies probe the region of the crystal that is penetrated by the excitation laser beam. Under resonance or near-resonance conditions, this is of the order of a few micrometres, meaning that there is a major benefit for an on-axis geometry such that the spectroscopically probed region corresponds to a part of the crystal that has been exposed to the X-ray beam during a crystallographic experiment.
We should note that we explicitly do not discuss the proper treatment of polarized
Raman or absorption spectra. As the crystal goniometer of our beamline at the Swiss
Light Source (SLS X10SA) at present has a single axis, it is not possible to re-orient
a crystal along a specific axis. Our micro-spectrophotometer is therefore at present
implemented without polarization analyzer stages. As a residual degree of polarization
owing to the fibre optics cannot be ruled out, for Raman spectra relative comparisons
of vibration-band amplitudes have to be undertaken with the crystal in a specific
orientation only (Carpentier et al., 2007).
2.2. Raman band sharpening and shifting
Raman spectra of crystals can exhibit several distinct differences compared with the
respective solution spectra of the same molecule. In solution, the molecules are subject
to constant movement, resulting in frequent intermolecular collisions and leading
to significant peak broadening. In contrast, the defined arrangement of molecules
in the ). This effect is even more pronounced if the sample is cooled, as is often the case
in macromolecular crystallography. On the other hand, the tight arrangement in the
crystal causes the intramolecular bond vibrations to be influenced by the neighbouring
molecules in the crystal, possibly resulting in peak shifting compared with the equivalent
solution spectra. Furthermore, vibrations in adjacent molecules in a crystal can either
be in phase or out of phase, possibly resulting in peak splitting or intensity fluctuations
owing to harmonic amplification or destruction (Hendra, 2002
). The direct consequence of this is that the in crystallo spectra can significantly differ from solution spectra and one has to be aware of
this during data interpretation.
2.3. Fluorescence background in Raman spectroscopy
The major practical drawback of Raman spectroscopy is the intrinsic background, usually
attributed to fluorescence, which can be up to 1011 and 103 times more intense than the actual nonresonance Raman or resonance Raman scattering,
respectively (Asher, 2001). This has long been a hindrance in the application of Raman spectroscopy as an effective
analytical tool in biology (Rippon et al., 1971
). In proteins it is particularly easy to excite unwanted fluorescence from, for example,
aromatic amino acids, buffers or cryoprotectants, or most commonly impurities and
contaminants. To minimize these effects, the protein sample has to be of the highest
possible purity before crystallization and, where possible, a careful choice of spectroscopically
silent crystallization reagents and cryoprotectants is beneficial. Another approach
to minimize the background fluorescence is to change the wavelength of the excitation
laser away from the absorption band of the fluorophore. Usually the background is
significantly reduced if excitation either in the UV or IR regions is utilized. Owing
to the complex and expensive instrumentation required, UV Raman measurements, while
extremely useful (Asher, 1988
; Kim et al., 2010
; Oladepo et al., 2012
), are significantly less common than IR Raman measurements. A typical laser wavelength
for nonresonance Raman experiments is 785 nm, as affordable high-power diode lasers
have become available and the sensitivity of CCD detectors is still high at these
wavelengths.
In the case of crystals and other solids it has been suggested that there are also
other influences on the measured background; for example, point defects on the surface
of crystals have been shown to contribute (Splett et al., 1997; Kim et al., 2010
). These intrinsic defects are unavoidable for protein crystals, and the resulting
background has to be dealt with either experimentally or during post-processing of
the spectroscopic data. Several methods have been suggested to reduce the Raman background.
The invasive method of photo-bleaching of the fluorophores (Macdonald & Wyeth, 2006
) is neither very effective nor desirable. Other methods include time-gating methods,
utilizing the fact that the Raman scattering lifetime is typically 10−11–10−13 s while the fluorescence lifetime is of the order of 10−6–10−9 s (Laubereau et al., 1972
; Van Duyne et al., 1974
). This method, however, is not applicable for in situ, in crystallo Raman measurements, since the required acquisition times are incompatible.
An especially useful technique for experimental background reduction is shifted excitation
Raman difference spectroscopy (SERDS; Shreve et al., 1992; Sowoidnich & Kronfeldt, 2012
). Here, two Raman spectra are recorded under identical conditions but utilizing two
different excitation lasers with wavelengths typically ∼0.5 nm apart. While the relatively
broad background is not affected by this small shift, the Raman peaks are shifted
exactly by the excitation wavelength difference. By differentiating the two spectra,
the background can be completely eliminated. The drawback of the method is the additional
investment in instrumentation and experimental time, as well as the difficult comparison
of the difference spectra with `normal' Raman spectra. On beamline X10SA (SLSpectroLAB)
at the Swiss Light Source spectroscopic facility we provide two integrated SERDS laser
systems at ∼785 and ∼647 nm and appropriate data-processing algorithms via the SLS-APE (SLS Laser Spectroscopy Analysis and Processing Environment) toolbox (see §
4.1).
3. Instrumentation
Accommodating a micro-spectrophotometer instrument into an existing MX experimental
beamline leads to compromising the maximal performance of the instrument. The sample
environment is usually crowded and occupied by various other devices such as a liquid-nitrogen
cryostat, an et al., 2014). The resulting micro-spectrophotometer, named MS3 (Dworkowski et al., in preparation) is based on the very successful on-axis MS2 device previously operated
in a mount-per-request mode (Pompidor et al., 2013
) and is always in place and available to beamline users. The on-axis geometry helps
to ensure good alignment of the X-ray beam and the spectroscopic light path (UV–visible
or Raman excitation laser) as well as to ensure that spectroscopic data are measured
from the X-ray-exposed volume of the crystal. A detailed description of the on-axis
micro-spectrophotometer is available elsewhere (Pompidor et al., 2013
).
4. Data analysis
As many of the previously mentioned interferences and data artefacts either cannot
be avoided or can only be avoided to a limited extent, careful post-processing of
the experimental data is imperative. To make this task easier and, more importantly,
more reproducible and reliable, we have developed a software toolbox for this kind
of data analysis. The SLS-APE toolbox is a MATLAB-based (The MathWorks Inc., Natick, USA) library of functions and data-handling procedures
for automated in crystallo data analysis. Raw data are loaded into a defined data object, which is then utilized
by all subprocedures. The toolbox is designed in such a way that the original raw
data are never unintentionally altered, but instead all derived data are added to
the object, allowing a certain level of backtracking. The basic structure of the data
object is shown in Fig. 3. Since the codebase is open source, it is very easy to expand the data object to
suit the experiment and analysis at hand. The toolbox is available from the corresponding
author upon request.
![]() |
Figure 3 The general structure of an SLS-APE data object. The raw data segment is never altered by the toolbox and all derived data can thus be tracked back to the original data set (pink). Additional data can be added depending on the experimental requirements (green). All data derived via the toolbox are assigned to dedicated data containers (beige). |
4.1. Capabilities of the toolbox
The main purpose of the development of the SLS-APE toolbox was the ability to analyze in crystallo spectroscopic data easily, flexibly and quantitatively. Special emphasis was put on making data comparison between different samples possible. Since most commercial spectrographs utilize a proprietary data format, the universal first step in the workflow is a conversion to the general SLS-APE data format, which is also readable by other MATLAB-based routines. At the SLSpectroLAB this entails a conversion from the spectrograph's proprietary software file format (ANDOR Solis, Andor Technology, Belfast, Northern Ireland), but the implementation of other formats is straightforward, in particular if a MATLAB module is supplied by the vendor. Depending on the type of spectroscopic data, this step is followed by calibration/conversion of the data. In the case of Raman data this is a calibration to the exact laser excitation wavelength and conversion to Raman shifts. For this purpose a user-friendly GUI is supplied with the toolbox, the Raman Calibration Tool (RaCaTo). A reference spectrum of acetaminophen, polystyrene, cyclohexane or silicon is used to calculate the exact laser wavelength, which is then in turn used to calibrate all subsequent spectra. This procedure, especially if repeated throughout the experiment, ensures exact spectral calibration even if the laser wavelength varies owing to aging, long runtimes or heating. Interpretation of convoluted or noisy spectra might pose a challenge. To aid the researcher in an initial assessment, the toolbox can perform basic peak-finding procedures as well as approximate band assignment to a standard or user-defined lookup table.
UV–visible absorption or fluorescence kinetic data usually do not require any additional
data processing, and the slicing routines of SLS-APE can be utilized directly to integrate peak areas and plot changes over time. In the
case of Raman spectra, however, a baseline correction is usually necessary. As discussed
previously, fluorescence background is a major problem for the weak Raman signal.
This background might be further increased if the experiment involves X-ray irradiation,
since the fluorescence can be enhanced by the radiation over time (McGeehan et al., 2007). This also means the background might not be constant and has to be determined and
corrected for each single spectrum in a kinetic series. Performing this task manually,
as is often performed in Raman experiments, is not only tedious but also extremely
error-prone and subject to user bias. Hence, an automated correction is necessary
for reliable data interpretation. Many algorithms for Raman background correction
have been suggested, including promising ones based on wavelets (Li, 2009
), multiple polynomials (Zhao et al., 2007
), `rolling ball' filtering (Kneen & Annegarn, 1996
), singular value decomposition (SVD; Palacký et al., 2011
) and principal component analysis (PCA; Hasegawa et al., 2000
). However, when evaluating these algorithms we found various problems, such as changing
backgrounds owing to an inability to deal with the varying fluorescence owing to X-ray
irradiation, lost peaks and negative values. The fastest and most reliable algorithm
we tested and therefore implemented in SLS-APE is the asymmetric least-squares (ALS) approach (Boelens et al., 2004
; Peng et al., 2010
), a finding supported by other systematic comparisons of correction models (Liland
et al., 2010
). Using the ALS algorithm allows fully automatic and reliable correction of very
strong and varying backgrounds so that usually no scaling within a kinetic series
is necessary. However, it is possible to scale data in SLS-APE, for example for the comparison of different samples or positions in a crystal. Further
processing depends on the experiment and can include difference spectra calculation,
time slices or dose-dependence calculations.
Another method to exclude background fluorescence from Raman spectra is a SERDS measurement.
This method does not rely on computational processing of the spectra but determines
the background experimentally. The drawbacks here are an effective doubling of the
acquisition time, which can be incompatible with kinetic measurements, and the fact
that the resulting spectra are difference spectra, making them difficult to compare
with non-SERDS reference spectra. To relieve this latter problem, SLS-APE offers the capability to convert SERDS spectra to a normal Raman signal using the
algorithm suggested by Matousek et al. (2005).
Finally, the toolbox provides mechanisms to extract, combine and export data for subsequent use in other applications.
4.2. Automation
An important aspect during the development of the SLS-APE toolbox was the possibility of automating the processing of repetitive or complex
measurements for convenience, reliability and speed. Hence, all of the tools utilize
the same basic data structure and input format. In combination with the powerful plotting
capabilities of MATLAB it is very easy to design complex analysis scripts. An example is the apefullshebang.m script for Raman spectra processing provided with the toolbox (Fig. 4). This script is used at the SLSpectroLab after each user shift to quickly convert
the data from the proprietary format used by the spectrograph into a universal format
(a free choice of comma-separated variable, tab-delimited or Microsoft Excel) to detect
and calibrate Raman data and to correct the baseline. In addition, the toolbox can
plot an overview sheet in PDF format containing all relevant information and spectra
for the user to take home for an initial overview of the acquired data (Fig. 5
). Owing to the quantity of data likely to be collected in a typical 16–24 h shift,
this overview can prove crucial in ensuring comprehensive and timely post-processing.
![]() |
Figure 4 Code example of the apefullshebang.m script used at the SLSpectroLAB. This script scans a complete folder of raw spectra and automatically processes them according to the experiment type. Finally, it outputs a PDF overview for the researcher as well as all generated data files in the chosen format. |
![]() |
Figure 5 Screenshot of the output of the SLS-APE script ramaninterleave.m. The raw Raman kinetic data (bottom) are automatically corrected for fluorescence background using the ALS algorithm (top). An invariant peak chosen by the user is then used to normalize the spectra (middle right) and the area under the peak of interest is integrated to plot peak decay against received dose in the subsequent step (middle left). All spectral intensities shown are in counts s−1. |
5. Case studies
Two case studies using Raman and absorption data acquired with the micro-spectrophotometer
at SLS beamline X10SA (Pompidor et al., 2013) will be used in the following section to demonstrate typical applications of the
SLS-APE toolbox.
5.1. Case study 1: X-ray energy dependence of S—S bond breakage in lysozyme
Radiation damage upon X-ray irradiation is a major problem in macromolecular crystallography.
Without cryocooling, crystalline samples degrade in the photon beam in a very short
time, apparent by loss of high-resolution diffraction spots and overall diffraction
power. This time is usually not long enough to measure a complete data set allowing
structure solution, although recent developments in high frame-rate detectors hold
the promise of a `room-temperature renaissance' (Owen et al., 2012, 2014
). However, even at cryogenic temperatures of around 100 K the sample will be affected
by the radiation, although at a much slower rate (Haas & Rossmann, 1970
; Hope, 1988
; Garman & Schneider, 1997
; Garman & Owen, 2006
). It is thus imperative for the structural biologist to know the extent of damage
the sample received during the diffraction experiment to ensure correct interpretation
of the structural model.
Disulfide bridges are commonly found in protein crystals and have been shown to be
very sensitive to radiation (Burmeister, 2000). To demonstrate the usefulness of the SLS-APE toolbox in quantifying the dose-dependence of disulfide-bond breakage at different
X-ray energies, we performed a stepwise radiation-exposure experiment on hen egg-white
lysozyme crystals. The crystals, with average dimensions of approximately 250 × 250
× 250 µm, were irradiated for a defined amount of time at fixed ω with an X-ray beam of 100 × 100 µm at 8.0, 12.4 and 15.0 keV, respectively, followed
by the acquisition of a non-resonance Raman spectrum with five accumulations of 20 s
acquisition time each. This was repeated with varying X-ray exposure times up to a
total dose of >15 MGy for each crystal and X-ray wavelength, allowing a comparison
to X-ray doses received by protein crystals during diffraction data collection, which
typically lie between 1 and 5 MGy. The resulting spectra were named according to the
pattern [spectrum name]_[t###][s###], where [t###] and [s###] denote the beam attenuation
and time of exposure, respectively. This information is not contained in the raw spectroscopic
file format, but by naming the files in this way the SLS-APE toolbox can be used for automated processing of nonstandard experiments. The files
are then processed with a simple script in SLS-APE consisting of the following steps: (i) conversion of each file to the SLS-APE format, (ii) calibration and conversion to Raman shift units, (iii) automatic baseline
correction and (iv) calculation and addition of the received dose per exposure by
calculation of the equivalent time of unattenuated beam irradiation multiplied by
the dose rate calculated using RADDOSE (Murray et al., 2004
; Paithankar et al., 2009
), (v) combination of all single data sets of one series into a single array of kinetic
data, (vi) integration of the in the region of interest (ROI), (vii) a plot of an overview of the raw and processed
data (Fig. 5
) and (viii) a plot of the normalized against the dose received (Fig. 6
).
![]() |
Figure 6 Plot of the decay of the 507 cm−1 normalized in a HEWL nonresonance Raman spectrum as a function of received X-ray dose at 8.0 keV (blue), 12.4 keV (red) and 15 keV (green). |
The resulting overlay of all three measurements shows identical dose dependences for
all three X-ray energies within a 95% confidence interval (see Supporting Information),
indicating that disulfide-bond breakage is energy-independent, as previously reported
(Shimizu et al., 2007). All three traces fit a previously proposed model in which back-conversion of the
anionic radical is significantly accelerated by X-rays, revealing an X-ray-induced
`repair' mechanism (Carpentier et al., 2010
).
The data and scripts discussed in this section will be available with the SLS-APE toolbox as a tutorial example.
5.2. Case study 2: X-ray of cytochrome c' from Shewanella frigidimarina
Cytochrome c′ from the marine microorganism S. frigidimarina is a haem-containing protein exhibiting pentacoordination of the haem iron with a
proximal histidine ligand (Manole et al., unpublished work). Crystals were grown over several days by the hanging-drop vapour-diffusion
method. 2 µl 20 mg ml−1 protein solution in 20 mM Tris–HCl pH 7 was mixed with an equivalent volume of reservoir solution consisting
of 0.1 M HEPES pH 7, 2.2 M ammonium sulfate. A UV–visible measured on a crystal prior to X-ray irradiation shows a (406 nm), broad peaks in the α/β region (460–580 nm) and a well defined charge-transfer band (638 nm) consistent with
the protein being predominantly in the ferric state (Fig. 7). The crystal was irradiated with X-rays of 12.4 keV at a dose rate of 15.4 kGy s−1 for 80 s, corresponding to a total absorbed dose of 1.23 MGy as calculated in RADDOSE (Murray et al., 2004
; Paithankar et al., 2009
), while a kinetic series of UV–visible absorption spectra was recorded. Each spectrum
was an accumulation of 20 exposures of 0.002 s, giving a total measurement interval
of 0.4 s with a corresponding dose per spectrum of 6.16 kGy. Kinetic data were exported
to the SLS-APE toolbox and analyzed using a simple script performing the following actions: (i)
conversion to the SLS-APE data format, (ii) smoothing of the spectra using a Savitzky–Golay algorithm (Fig.
7
, second row), (iii) peak identification in the first and last spectrum of the series
(Fig. 7
, third row) and (iv) kinetic time-slices through the changing peaks (Fig. 7
, fourth row). The results are plotted and the derived data as well as the generated
figures are automatically saved to the user's directory.
![]() |
Figure 7 Processing of kinetic data for the X-ray reduction of cytochrome c′ from S. frigidimarina. The panels in the second row show the smoothed data, those in the third row show the peaks identified in the first and the last spectrum and those in the fourth row show kinetic slices through those peaks. The spectrum of the ferric protein (blue) is rapidly interconverted to that of the ferrous form. Bottom panel: the dose-dependence of absorbance at three spectral peaks. Changes are largely complete after an absorbed dose of 1 MGy. |
Upon irradiation, the spectrum corresponding to the ferric form of the protein showed rapid interconversion to a spectrum consistent with the ferrous form of the protein. The α/β region as well as loss of the charge-transfer peak could be observed. Reduction appeared to be complete after an accumulated dose of 1 MGy.
shifted to a split peak (429 and 441 nm) and sharpening of the peaks in the6. Conclusion and future studies
The SLS-APE toolbox has been shown to be effective in aiding researchers in processing, background subtraction and displaying single-crystal spectroscopic data. The toolbox may be applied to UV–visible absorption, fluorescence and both resonance Raman and nonresonance Raman experiments. The application of the best-practice background-subtraction algorithms, together with peak identification and plotting, makes it easier for users to understand the spectroscopic data measured. The high-throughput and automated nature of the toolbox makes it particularly useful in kinetic or otherwise complex experiments where many spectra must be processed in an efficient, reproducible and consistent way. As the current version of the toolbox resulted from a natural evolution of individual analysis tools, it can be optimized for general use.
Acknowledgements
We would like to thank the X10SA beamline partners for funding: the Max Planck Society (MPG) and the pharmaceutical companies Novartis and F. Hoffmann-La Roche. We would also like to thank our scientific collaborators Hans-Petter Hersleth and Åsmund K. Røhr at the University of Oslo, Dominique Bourgeois at Institut de Biologie Structurale Grenoble and Antonello Merlino and Alessandro Vergara at the University of Naples for continuous input, feedback and feature requests, which led to the development of the SLS-APE toolbox. We acknowledge the assistance of Andreea A. Manole at the University of Essex in the preparation of crystals of S. frigidimarina cytochrome c′. Part of this work was carried out under SLS long-term beamtime award 20111166 and was funded in part by EU FP7 BioStructX award 2370.
References
Adam, V., Carpentier, P., Violot, S., Lelimousin, M., Darnault, C., Nienhaus, G. U.
& Bourgeois, D. (2009). J. Am. Chem. Soc. 131, 18063–18065. Web of Science CrossRef PubMed CAS Google Scholar
Allan, E. G., Kander, M. C., Carmichael, I. & Garman, E. F. (2013). J. Synchrotron Rad. 20, 23–36. Web of Science CrossRef CAS IUCr Journals Google Scholar
Asher, S. A. (1988). Annu. Rev. Phys. Chem. 39, 537–588. CrossRef CAS PubMed Google Scholar
Asher, S. A. (2001). Handbook of Vibrational Spectroscopy, pp. 557–571. New York: John Wiley & Sons. Google Scholar
Beitlich, T., Kühnel, K., Schulze-Briese, C., Shoeman, R. L. & Schlichting, I. (2007).
J. Synchrotron Rad. 14, 11–23. Web of Science CrossRef CAS IUCr Journals Google Scholar
Boelens, H. F. M., Dijkstra, R. J., Eilers, P. H. C., Fitzpatrick, F. & Westerhuis,
J. A. (2004). J. Chromatogr. A, 1057, 21–30. Web of Science CrossRef PubMed CAS Google Scholar
Burmeister, W. P. (2000). Acta Cryst. D56, 328–341. Web of Science CrossRef CAS IUCr Journals Google Scholar
Carey, P. R. (1978). Q. Rev. Biophys. 11, 309–370. CrossRef CAS PubMed Google Scholar
Carey, P. R. (1999). J. Biol. Chem. 274, 26625–26628. Web of Science CrossRef PubMed CAS Google Scholar
Carpentier, P., Royant, A., Ohana, J. & Bourgeois, D. (2007). J. Appl. Cryst. 40, 1113–1122. Web of Science CrossRef CAS IUCr Journals Google Scholar
Carpentier, P., Royant, A., Weik, M. & Bourgeois, D. (2010). Structure, 18, 1410–1419. Web of Science CrossRef CAS PubMed Google Scholar
Davies, R. J., Burghammer, M. & Riekel, C. (2009). J. Synchrotron Rad. 16, 22–29. Web of Science CrossRef CAS IUCr Journals Google Scholar
Diamond, R. (1974). J. Mol. Biol. 82, 371–391. CrossRef CAS PubMed Web of Science Google Scholar
Ellis, M. J., Buffey, S. G., Hough, M. A. & Hasnain, S. S. (2008). J. Synchrotron Rad. 15, 433–439. Web of Science CrossRef CAS IUCr Journals Google Scholar
Fuchs, M. R. et al. (2014). J. Synchrotron Rad. 21, 340–351. Web of Science CrossRef CAS IUCr Journals Google Scholar
Garman, E. F. & Owen, R. L. (2006). Acta Cryst. D62, 32–47. Web of Science CrossRef CAS IUCr Journals Google Scholar
Garman, E. F. & Schneider, T. R. (1997). J. Appl. Cryst. 30, 211–237. CrossRef Web of Science IUCr Journals Google Scholar
Garman, E. F. & Weik, M. (2011). J. Synchrotron Rad. 18, 313–317. Web of Science CrossRef IUCr Journals Google Scholar
Gouadec, G. & Colomban, P. (2007). Prog. Cryst. Growth Charact. Mater. 53, 1–56. Web of Science CrossRef CAS Google Scholar
Haas, D. J. & Rossmann, M. G. (1970). Acta Cryst. B26, 998–1004. CrossRef CAS IUCr Journals Web of Science Google Scholar
Harrand, M. & Lennuier, R. (1946). C. R. Hebd. Acad. Sci. 223, 356–357. CAS Google Scholar
Hasegawa, T., Nishijo, J. & Umemura, J. (2000). Chem. Phys. Lett. 317, 642–646. Web of Science CrossRef CAS Google Scholar
He, C., Fuchs, M. R., Ogata, H. & Knipp, M. (2012). Angew. Chem. Int. Ed. 51, 4470–4473. Web of Science CrossRef CAS Google Scholar
Hendra, P. J. (2002). Internet J. Vibr. Spectrosc. 6, 3. Google Scholar
Hersleth, H.-P. & Andersson, K. K. (2011). Biochim. Biophys. Acta, 1814, 785–796. Web of Science CrossRef CAS PubMed Google Scholar
Hope, H. (1988). Acta Cryst. B44, 22–26. CrossRef CAS Web of Science IUCr Journals Google Scholar
Hough, M. A., Antonyuk, S. V., Strange, R. W., Eady, R. R. & Hasnain, S. S. (2008).
J. Mol. Biol. 378, 353–361. Web of Science CrossRef PubMed CAS Google Scholar
Julien, C. (1980). J. Opt. 11, 257–267. CrossRef CAS Web of Science Google Scholar
Kim, H., Kosuda, K. M., Van Duyne, R. P. & Stair, P. C. (2010). Chem. Soc. Rev. 39, 4820–4844. Web of Science CrossRef CAS PubMed Google Scholar
Kneen, M. A. & Annegarn, H. J. (1996). Nucl. Instrum. Methods Phys. Res. B, 109/110, 209–213. CrossRef CAS Web of Science Google Scholar
Laubereau, A., von der Linde, D. & Kaiser, W. (1972). Phys. Rev. Lett. 28, 1162–1165. CrossRef CAS Web of Science Google Scholar
Li, G. (2009). 2009 ETP International Conference on Future Computer and Communication, pp. 198–200. https://doi.ieeecomputersociety.org/10.1109/FCC.2009.69 Google Scholar
Liland, K. H., Almøy, T. & Mevik, B. H. (2010). Appl. Spectrosc. 64, 1007–1016. CrossRef CAS PubMed Google Scholar
Macdonald, A. M. & Wyeth, P. (2006). J. Raman Spectrosc. 37, 830–835. Web of Science CrossRef CAS Google Scholar
Matousek, P., Towrie, M. & Parker, A. W. (2005). Appl. Spectrosc. 59, 848–851. Web of Science CrossRef PubMed CAS Google Scholar
McGeehan, J. E., Carpentier, P., Royant, A., Bourgeois, D. & Ravelli, R. B. G. (2007).
J. Synchrotron Rad. 14, 99–108. Web of Science CrossRef CAS IUCr Journals Google Scholar
Meents, A., Owen, R. L., Murgida, D., Hildebrandt, P., Schneider, R., Pradervand,
C., Bohler, P. & Schulze-Briese, C. (2007). AIP Conf. Proc. 879, 1984–1988. CrossRef CAS Google Scholar
Merlino, A., Fuchs, M. R., Pica, A., Balsamo, A., Dworkowski, F. S. N., Pompidor,
G., Mazzarella, L. & Vergara, A. (2013). Acta Cryst. D69, 137–140. Web of Science CrossRef CAS IUCr Journals Google Scholar
Murray, J. W., Garman, E. F. & Ravelli, R. B. G. (2004). J. Appl. Cryst. 37, 513–522. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ohana, I., Yacoby, Y. & Bezalel, M. (1986). Rev. Sci. Instrum. 57, 9. CrossRef Web of Science Google Scholar
Oladepo, S. A., Xiong, K., Hong, Z., Asher, S. A., Handen, J. & Lednev, I. K. (2012).
Chem. Rev. 112, 2604–2628. Web of Science CrossRef CAS PubMed Google Scholar
Orville, A. M., Buono, R., Cowan, M., Héroux, A., Shea-McCarthy, G., Schneider, D.
K., Skinner, J. M., Skinner, M. J., Stoner-Ma, D. & Sweet, R. M. (2011). J. Synchrotron Rad. 18, 358–366. Web of Science CrossRef CAS IUCr Journals Google Scholar
Owen, R. L., Axford, D., Nettleship, J. E., Owens, R. J., Robinson, J. I., Morgan,
A. W., Doré, A. S., Lebon, G., Tate, C. G., Fry, E. E., Ren, J., Stuart, D. I. & Evans,
G. (2012). Acta Cryst. D68, 810–818. Web of Science CrossRef CAS IUCr Journals Google Scholar
Owen, R. L., Paterson, N., Axford, D., Aishima, J., Schulze-Briese, C., Ren, J., Fry,
E. E., Stuart, D. I. & Evans, G. (2014). Acta Cryst. D70, 1248–1256. Web of Science CrossRef IUCr Journals Google Scholar
Owen, R. L., Pearson, A. R., Meents, A., Boehler, P., Thominet, V. & Schulze-Briese,
C. (2009). J. Synchrotron Rad. 16, 173–182. Web of Science CrossRef CAS IUCr Journals Google Scholar
Paithankar, K. S., Owen, R. L. & Garman, E. F. (2009). J. Synchrotron Rad. 16, 152–162. Web of Science CrossRef CAS IUCr Journals Google Scholar
Palacký, J., Mojzeš, P. & Bok, J. (2011). J. Raman Spectrosc. 42, 1528–1539. Google Scholar
Palings, I., Pardoen, J. A., van den Berg, E., Winkel, C., Lugtenburg, J. & Mathies,
R. A. (1987). Biochemistry, 26, 2544–2556. CrossRef CAS PubMed Web of Science Google Scholar
Pearson, A. R., Pahl, R., Kovaleva, E. G., Davidson, V. L. & Wilmot, C. M. (2007).
J. Synchrotron Rad. 14, 92–98. Web of Science CrossRef CAS IUCr Journals Google Scholar
Peng, J., Peng, S., Jiang, A., Wei, J., Li, C. & Tan, J. (2010). Anal. Chim. Acta, 683, 63–68. Web of Science CrossRef CAS PubMed Google Scholar
Pompidor, G., Dworkowski, F. S. N., Thominet, V., Schulze-Briese, C. & Fuchs, M. R.
(2013). J. Synchrotron Rad. 20, 765–776. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rajendran, C., Dworkowski, F. S. N., Wang, M. & Schulze-Briese, C. (2011). J. Synchrotron Rad. 18, 318–328. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rippon, W. B., Koenig, J. L. & Walton, A. G. (1971). J. Agric. Food Chem. 19, 692–697. CrossRef CAS PubMed Web of Science Google Scholar
Rossi, G. L. & Bernhard, S. A. (1970). J. Mol. Biol. 49, 85–91. CrossRef CAS PubMed Web of Science Google Scholar
Royant, A., Carpentier, P., Ohana, J., McGeehan, J., Paetzold, B., Noirclerc-Savoye,
M., Vernède, X., Adam, V. & Bourgeois, D. (2007). J. Appl. Cryst. 40, 1105–1112. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sakai, K., Matsui, Y., Kouyama, T., Shiro, Y. & Adachi, S. (2002). J. Appl. Cryst. 35, 270–273. Web of Science CrossRef CAS IUCr Journals Google Scholar
Schlichting, I., Berendzen, J., Chu, K., Stock, A. M., Maves, S. A., Benson, D. E.,
Sweet, R. M., Ringe, D., Petsko, G. A. & Sligar, S. G. (2000). Science, 287, 1615–1622. Web of Science CrossRef PubMed CAS Google Scholar
Shimizu, N., Hirata, K., Hasegawa, K., Ueno, G. & Yamamoto, M. (2007). J. Synchrotron Rad. 14, 4–10. Web of Science CrossRef CAS IUCr Journals Google Scholar
Shimizu, N., Shimizu, T., Baba, S., Hasegawa, K., Yamamoto, M. & Kumasaka, T. (2013).
J. Synchrotron Rad. 20, 948–952. Web of Science CrossRef CAS IUCr Journals Google Scholar
Shreve, A. P., Cherepy, N. J. & Mathies, R. A. (1992). Appl. Spectrosc. 46, 707–711. CrossRef CAS Web of Science Google Scholar
Sowoidnich, K. & Kronfeldt, H.-D. (2012). ISRN Spectrosc. 2012, 256326. Google Scholar
Spiro, T. G. & Strekas, T. C. (1974). J. Am. Chem. Soc. 96, 338–345. CrossRef CAS PubMed Web of Science Google Scholar
Splett, A., Splett, C. & Pilz, W. (1997). J. Raman Spectrosc. 28, 481–485. CrossRef CAS Google Scholar
Stoner-Ma, D., Skinner, J. M., Schneider, D. K., Cowan, M., Sweet, R. M. & Orville,
A. M. (2011). J. Synchrotron Rad. 18, 37–40. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sutton, K. A., Black, P. J., Mercer, K. R., Garman, E. F., Owen, R. L., Snell, E.
H. & Bernhard, W. A. (2013). Acta Cryst. D69, 2381–2394. Web of Science CrossRef CAS IUCr Journals Google Scholar
Thomas, G. J. (1999). Annu. Rev. Biophys. Biomol. Struct. 28, 1–27. Web of Science CrossRef PubMed CAS Google Scholar
Van Duyne, R. P., Jeanmaire, D. L. & Shriver, D. F. (1974). Anal. Chem. 46, 213–222. CrossRef CAS Web of Science Google Scholar
Zhao, J., Lui, H., McLean, D. I. & Zeng, H. (2007). Appl. Spectrosc. 61, 1225–1232. Web of Science CrossRef PubMed CAS Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.