CORRIGENDUM: Fixing the MRI R2-iron calibration in liver.
Doyle, E., Ghugre, N., Coates, T.D. and Wood, J.C. (2020), Fixing the MRI R2-iron calibration in liver. Am J Hematol, 95: E120-E122. doi:10.1002/ajh.25754
CORRIGENDUM: The original article which appeared in Am J Hematol, 95: E120-E122. contains one typographical error in Equation # 2, which appeared as LIC=0.15*(R2)1.3555 instead of the correct formulation LIC=0.015*(R2)1.3555.
The authors would like to thank Kartik Jhaveri, Gerard Healy, Stephan Kannengiesser, and Arthur Wunderlich for calling this error to our attention and apologize for any confusion regarding this point.