Volume 22, Issue 6 pp. 503-510
Free to Read

Impact of Occluder Device Type on Success of Percutaneous Closure of Atrial Septal Defects–A Medium-Term Follow-up Study

MICHAEL BECKER M.D.

MICHAEL BECKER M.D.

University Hospital RWTH Aachen, Department of Cardiology, Aachen, Germany

RWTH Aachen University, Chair of Medical Engineering, Aachen, Germany

Both authors contributed equally.

Search for more papers by this author
DOROTHEE FRINGS

DOROTHEE FRINGS

University Hospital RWTH Aachen, Department of Cardiology, Aachen, Germany

Both authors contributed equally.

Search for more papers by this author
JÖRG SCHRÖDER M.D.

JÖRG SCHRÖDER M.D.

University Hospital RWTH Aachen, Department of Cardiology, Aachen, Germany

Search for more papers by this author
CHRISTINA OCKLENBURG M.Sc.

CHRISTINA OCKLENBURG M.Sc.

University Hospital RWTH Aachen, Department of Medical Statistics, Aachen, Germany

Search for more papers by this author
EBERHARD MÜHLER M.D.

EBERHARD MÜHLER M.D.

University Hospital RWTH Aachen, Department of Pediatric Cardiology, Aachen, Germany

Search for more papers by this author
RAINER HOFFMANN M.D.

RAINER HOFFMANN M.D.

University Hospital RWTH Aachen, Department of Cardiology, Aachen, Germany

Search for more papers by this author
ANDREAS FRANKE M.D.

ANDREAS FRANKE M.D.

Klinikum Siloah, Department of Cardiology, Hannover, Germany

Search for more papers by this author
WOLFGANG LEPPER M.D.

WOLFGANG LEPPER M.D.

University Hospital RWTH Aachen, Department of Cardiology, Aachen, Germany

Search for more papers by this author
First published: 01 December 2009
Citations: 9
Address for reprints: Michael Becker, M.D., Medical Clinic I, University RWTH Aachen, Pauwelsstraße 30, 52057 Aachen, Germany. Fax: +49-241-8082303; e-mail: [email protected]

Abstract

Objective: To identify differences between various occluder types regarding thromboembolic event rates and existence of residual shunts during medium-term follow-up.

Methods: Three hundred nine consecutive patients with presumed paradoxical embolism and scheduled for percutaneous closure of patent foramen ovale or atrial septal defect between 1997 and 2006 were considered for this study. Device implantation failed in seven patients (nonstable device position), so 302 patients (46 ± 12 years, 190 males) formed the study group using Amplatzer™ (n = 208), Starflex™ (n = 61), and Cardiastar™ (n = 33) occluders. Follow-up transesophageal echocardiography was performed 1, 3, and 6 months after implantation.

Results: Periinterventional complications occurred in two patients (cardiac arrhythmias). There were more residual shunts in the Starflex™ and Cardiastar™ group than in the Amplatzer™ group at 6-month follow-up (8/61 vs. 7/33 vs. 8/208, P = 0.0005). Performing logistic regression, the type of occluder device was a significant risk factor for the presence of residual shunt 6 months after implantation (P = 0.0033; Cardiastar™ vs. Amplatzer™ OR 6.346, 95% CI 1.998 – 20.156; Starflex™ vs. Amplatzer™ OR 4.369, 95% CI 1.444 – 13.222). During mean follow-up of 31 ± 8 months 16 recurrent thromboembolic events occurred; the annual recurrence was 2.1% for combined end-point stroke, TIA, and peripheral embolism (Amplatzer™: 1.3%, Starflex™: 3.9%, and Cardiastar™: 3.6%, P = 0.0467). The presence of an atrial septal aneurysm was the only significant risk factor (P = 0.0168, OR 3.664, 95% CI 1.263 – 10.630) for the occurrence of thromboembolic events.

Conclusions: Percutaneous closure of patent foramen ovale or atrial septal defect is a safe procedure with little incidence of peri- and postprocedural complications. There is a significant difference between the Amplatzer™, Cardiastar™, and Starflex™ occluders in regard to complete closure of the defects and annual recurrence of thromboembolic events.

The full text of this article hosted at iucr.org is unavailable due to technical difficulties.