Computational Design and In Vitro Characterization of an Integrated Maglev Pump-Oxygenator

Juntao Zhang

Juntao Zhang

Artificial Organs Laboratory, Department of Surgery, University of Maryland School of Medicine, Baltimore, MD;

Search for more papers by this author
M. Ertan Taskin

M. Ertan Taskin

Artificial Organs Laboratory, Department of Surgery, University of Maryland School of Medicine, Baltimore, MD;

Search for more papers by this author
Andrew Koert

Andrew Koert

Levitronix LLC, Waltham; and

Search for more papers by this author
Tao Zhang

Tao Zhang

Artificial Organs Laboratory, Department of Surgery, University of Maryland School of Medicine, Baltimore, MD;

Search for more papers by this author
Barry Gellman

Barry Gellman

Levitronix LLC, Waltham; and

Search for more papers by this author
Kurt A. Dasse

Kurt A. Dasse

Levitronix LLC, Waltham; and

Search for more papers by this author
Richard J. Gilbert

Richard J. Gilbert

Department of Mechanical Engineering, Massachusetts Institute of Technology, Cambridge, MA, USA

Search for more papers by this author
Bartley P. Griffith

Bartley P. Griffith

Artificial Organs Laboratory, Department of Surgery, University of Maryland School of Medicine, Baltimore, MD;

Search for more papers by this author
Zhongjun J. Wu

Corresponding Author

Zhongjun J. Wu

Artificial Organs Laboratory, Department of Surgery, University of Maryland School of Medicine, Baltimore, MD;

Dr. Zhongjun J. Wu, Department of Surgery, University of Maryland School of Medicine, MSTF-436, 10. S. Pine Street, Baltimore, MD 21201, USA. E-mail: [email protected]Search for more papers by this author
First published: 20 October 2009
Citations: 29

Abstract

For the need for respiratory support for patients with acute or chronic lung diseases to be addressed, a novel integrated maglev pump-oxygenator (IMPO) is being developed as a respiratory assist device. IMPO was conceptualized to combine a magnetically levitated pump/rotor with uniquely configured hollow fiber membranes to create an assembly-free, ultracompact system. IMPO is a self-contained blood pump and oxygenator assembly to enable rapid deployment for patients requiring respiratory support or circulatory support. In this study, computational fluid dynamics (CFD) and computer-aided design were conducted to design and optimize the hemodynamics, gas transfer, and hemocompatibility performances of this novel device. In parallel, in vitro experiments including hydrodynamic, gas transfer, and hemolysis measurements were conducted to evaluate the performance of IMPO. Computational results from CFD analysis were compared with experimental data collected from in vitro evaluation of the IMPO. The CFD simulation demonstrated a well-behaved and streamlined flow field in the main components of this device. The results of hydrodynamic performance, oxygen transfer, and hemolysis predicted by computational simulation, along with the in vitro experimental data, indicate that this pump-lung device can provide the total respiratory need of an adult with lung failure, with a low hemolysis rate at the targeted operating condition. These detailed CFD designs and analyses can provide valuable guidance for further optimization of this IMPO for long-term use.