Volume 69, Issue 3 pp. 862-867
Full Paper

On shimming approaches in 3T breast MRI

Ileana Hancu

Corresponding Author

Ileana Hancu

GE Global Research Center, Niskayuna, New York, USA

GE Global Research Center, One Research Circle, K1-NMR133, Niskayuna, NY 12309===Search for more papers by this author
Ambey Govenkar

Ambey Govenkar

Extenprise Inc., Pune, Maharashtra, India

Search for more papers by this author
Robert E. Lenkinski

Robert E. Lenkinski

Department of Radiology, UT Southwestern Medical Center, Dallas, Texas, USA

Search for more papers by this author
Seung-Kyun Lee

Seung-Kyun Lee

GE Global Research Center, Niskayuna, New York, USA

Search for more papers by this author
First published: 03 May 2012
Citations: 27

Abstract

A comparative study is presented, analyzing quantitatively the impact of 15 shim strategies on the homogeneity of the main magnetic field over the three-dimensional breast region in 3T MRI. The results obtained in 12 female volunteers, spanning a wide range of body and breast types, indicate that the inclusion of the back and heart in the shim region of interest leads to considerable decrease in field homogeneity, and needs to be avoided. Comparison between shim strategies using volumetric B0 maps, covering the entire breast region, and 1–6 plane B0 maps indicate only minimally reduced performance for the latter. Interestingly, however, no single shim strategy relying on a limited number of B0 maps as input was found to work best in all volunteers. This was attributed to the limited capability of a small number of B0 maps to capture the B0 variability existent within breast. On the average, a rectangular shim region of interest, encompassing the breast region alone, worked best for the cohort studied here. Magn Reson Med, 2013. © 2012 Wiley Periodicals, Inc.

The full text of this article hosted at iucr.org is unavailable due to technical difficulties.