Volume 34, Issue 3 pp. 494-497
Notes
Full Access

Quite transverse gradiant coils: Lorentz force balanced designs using geometrical similitude

Richard W. Bowtell

Richard W. Bowtell

Magnetic Resonance Centre, University of Nottingham, Notting-ham, United Kingdom.

Search for more papers by this author
Peter Mansfield FRS

Corresponding Author

Peter Mansfield FRS

Magnetic Resonance Centre, University of Nottingham, Notting-ham, United Kingdom.

Magnetic Resonance Centre, University of Nottingham, University Park, Nottingham, NG72RD, UK===Search for more papers by this author
First published: September 1995
Citations: 39

Abstract

The principle of active acoustic screening, recently introduced by Mansfield et al. (1) to reduce acoustic noise ingradient coils for MRI, is applied to the design of transverse distributed gradient coils of the fingerprint design. The results indicate that for complete Lorentz force balancing and for exact satisfaction of Kirchoff's law, it is necessary to have a coil system wound on four cylinders if full active magnetic screening of the assembly is required.