Evaluation of a new methodology to simulate damage and wear of polyethylene hip replacements subjected to edge loading in hip simulator testing
Susan Partridge
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
Search for more papers by this authorJoanne L. Tipper
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
Search for more papers by this authorMazen Al-Hajjar
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
Search for more papers by this authorGraham H. Isaac
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
DePuy Synthes Joint Reconstruction, Leeds, United Kingdom
Search for more papers by this authorJohn Fisher
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
Search for more papers by this authorCorresponding Author
Sophie Williams
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
Correspondence to: S. Williams; e-mail: [email protected]Search for more papers by this authorSusan Partridge
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
Search for more papers by this authorJoanne L. Tipper
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
Search for more papers by this authorMazen Al-Hajjar
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
Search for more papers by this authorGraham H. Isaac
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
DePuy Synthes Joint Reconstruction, Leeds, United Kingdom
Search for more papers by this authorJohn Fisher
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
Search for more papers by this authorCorresponding Author
Sophie Williams
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom
Correspondence to: S. Williams; e-mail: [email protected]Search for more papers by this authorAbstract
Wear and fatigue of polyethylene acetabular cups have been reported to play a role in the failure of total hip replacements. Hip simulator testing under a wide range of clinically relevant loading conditions is important. Edge loading of hip replacements can occur following impingement under extreme activities and can also occur during normal gait, where there is an offset deficiency and/or joint laxity. This study evaluated a hip simulator method that assessed wear and damage in polyethylene acetabular liners that were subjected to edge loading. The liners tested to evaluate the method were a currently manufactured crosslinked polyethylene acetabular liner and an aged conventional polyethylene acetabular liner. The acetabular liners were tested for 5 million standard walking cycles and following this 5 million walking cycles with edge loading. Edge loading conditions represented a separation of the centers of rotation of the femoral head and the acetabular liner during the swing phase, leading to loading of the liner rim on heel strike. Rim damage and cracking was observed in the aged conventional polyethylene liner. Steady-state wear rates assessed gravimetrically were lower under edge loading compared to standard loading. This study supports previous clinical findings that edge loading may cause rim cracking in liners, where component positioning is suboptimal or where material degradation is present. The simulation method developed has the potential to be used in the future to test the effect of aging and different levels of severity of edge loading on a range of cross-linked polyethylene materials. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 1456–1462, 2018.
REFERENCES
- 1National Joint Registry. 12th Annual Report. 2014;
- 2 Tipper JL, Ingham E, Hailey JL, Besong AA, Fisher J, Wroblewski BM, Stone MH. Quantitative analysis of polyethylene wear debris, wear rate and head damage in retrieved Charnley hip prostheses. J Mater Sci Mater Med 2000; 11: 117–124.
- 3 Ingham E, Fisher J. Biological reactions to wear debris in total joint replacement. Proc Inst Mech Eng H 2000; 214: 21–37.
- 4 Revell PA, Al-Saffar N, Kobayashi A. Biological reaction to debris in relation to joint prostheses. Proc Inst Mech Eng H 1997; 211: 187–197.
- 5 Sutula LC, Collier JP, Saum KA, Currier BH, Currier JH, Sanford WM, Mayor MB, Wooding RE, Sperling DK, Williams IR. The Otto Aufranc Award. Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin Orthop Relat Res 1995: 28– 40.
- 6 Currier BH, Currier JH, Mayor MB, Lyford KA, Van Citters DW, Collier JP. In vivo oxidation of γ-barrier-sterilized ultra-high-molecular-weight polyethylene bearings. J Arthroplasty 2007; 22: 721–731.
- 7 Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg Am 2007; 89: 2023–2029.
- 8 Kurtz SM, Hozack WJ, Purtill JJ, Marcolongo M, Kraay MJ, Goldberg VM, Sharkey PF, Parvizi J, Rimnac CM, Edidin AA. Significance of in vivo degradation for polyethylene in total hip arthroplasty. J Clin Orthop Relat Res 2006; 453: 47–57.
- 9 McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res 1999; 17: 157–167.
- 10 Muratoglu OK, Bragdon CR, O'Connor DO, Jasty M, Harris WH, Rizwan G, McGarry F. Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE). Biomaterials 1999; 20: 1463–1470.
- 11
Baker DA,
Hastings RS,
Pruitt L. Study of fatigue resistance of chemical and radiation crosslinked medical grade ultrahigh molecular weight polyethylene. J Biomed Mater Res 1999; 46: 573–581.
10.1002/(SICI)1097-4636(19990915)46:4<573::AID-JBM16>3.0.CO;2-A CAS PubMed Web of Science® Google Scholar
- 12 Furmanski J, Anderson M, Bal S, Greenwald AS, Halley D, Penenberg B, Ries M, Pruitt L. Clinical fracture of cross-linked UHMWPE acetabular liners. Biomaterials Elsevier Ltd; 2009; 30: 5572–5582.
- 13 Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am 2007; 89: 2212–2217.
- 14 Birman M V., Noble PC, Conditt MA, Li S, Mathis KB. Cracking and impingement in ultra-high-molecular-weight polyethylene acetabular liners. J Arthroplasty 2005; 20: 87–92.
- 15 Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty 1996; 11: 377–389.
- 16
Fisher J,
Al-Hajjar M,
Williams S,
Jennings LM,
Ingham E. In vitro measurement of wear in joint replacements: A stratified approach for enhanced reliability “safer” pre-clinical simulation testing. Semin Arthroplasty Elsevier; 2012; 23: 286–288.
10.1053/j.sart.2012.12.003 Google Scholar
- 17BS ISO 14242:2002. Implants for Surgery — Wear of Total Hip-Joint Prostheses. 2002.
- 18 Bergmann G, Graichen F, Rohlmann A. Is staircase walking a risk for the fixation of hip implants? J Biomech 1995; 28: 535–553.
- 19 Bergmann G, Graichen F, Rohlmann A. Hip joint contact forces during stumbling. Langenbeck's Arch Surg 2004; 389: 53–59.
- 20 Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA. Impingement in total hip arthroplasty. J Arthroplasty 2005; 20: 427–435.
- 21 Usrey MM, Noble PC, Rudner LJ, Conditt MA, Birman MV, Santore RF, Mathis KB. Does neck/liner impingement increase wear of ultrahigh-molecular-weight polyethylene liners? J Arthroplasty. 2006; 21: 65–71.
- 22
Duffy GP,
Wannomae KK,
Rowell SL,
Muratoglu OK. Fracture of a cross-linked polyethylene liner due to impingement. J Arthroplasty Elsevier Inc; 2009; 24: 158.e15–158.e19.
10.1016/j.arth.2007.12.020 Google Scholar
- 23 Marchetti E, Krantz N, Berton C, Bocquet D, Fouilleron N, Migaud H, Girard J. Component impingement in total hip arthroplasty: Frequency and risk factors. A continuous retrieval analysis series of 416 cup. Orthop Traumatol Surg Res Elsevier Masson SAS; 2011; 97: 127–133.
- 24 Nevelos JE, Ingham E, Doyle C, Fisher J, Nevelos AB. Analysis of retrieved alumina ceramic components from Mittelmeier total hip prostheses. Biomaterials 1999; 20: 1833–1840.
- 25 Nevelos J, Ingham E, Doyle C, Streicher R, Nevelos A, Walter W, Fisher J. Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplasty 2000; 15: 793–795.
- 26 Al-Hajjar M, Jennings LM, Begand S, Oberbach T, Delfosse D, Fisher J. Wear of novel ceramic-on-ceramic bearings under adverse and clinically relevant hip simulator conditions. J Biomed Mater Res B Appl Biomater 2013; 101: 1456–1462.
- 27 Al-Hajjar M, Fisher J, Tipper JL, Williams S, Jennings LM. Wear of 36-mm BIOLOX (R) delta ceramic-on-ceramic bearing in total hip replacements under edge loading conditions. Proc Inst Mech Eng H 2013; 227: 535–542.
- 28 Al-Hajjar M, Leslie IJ, Tipper J, Williams S, Fisher J, Jennings LM. Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX(R) delta ceramic-on-ceramic total hip replacement. J Biomed Mater Res B Appl Biomater 2010; 95: 263–268.
- 29 Leslie IJ, Williams S, Isaac G, Ingham E, Fisher J. High cup angle and microseparation increase the wear of hip surface replacements. Clin Orthop Relat Res 2009; 467: 2259–2265.
- 30 Dennis DA, Komistek RD, Northcut EJ, Ochoa JA, Ritchie A. “In vivo” determination of hip joint separation and the forces generated due to impact loading conditions. J Biomech 2001; 34: 623–629.
- 31 Komistek RD, Dennis DA, Ochoa JA, Haas BD, Hammill C. In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty. J Bone Joint Surg Am 2002; 84–A: 1836–1841.
- 32 Barbour PS, Stone MH, Fisher J. A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates. Proc Inst Mech Eng H 1999; 213: 455–467.
- 33 Schroder DT, Kelly NH, Wright TM, Parks ML. Retrieved highly crosslinked UHMWPE acetabular liners have similar wear damage as conventional UHMWPE. Clin Orthop Relat Res 2011; 469: 387–394.
- 34 Bell CJ, Wlaker PS, Aberysundra MR, Simmons JM, King PM, Blunn GW. Effect of oxidation on delamination of ultrahigh-molecular-weight polyethylene tibial components. J Arthroplasty 1998; 13: 280–290.
- 35 Blunn G, Joshi AB, Lilley PA, Engelbrecht E, Ryd L, Lidgren L, Hardinge K, Nieder E, Walker PS. Polyethylene wear in unicondylar knee prostheses. 106 retrieved Marmor, PCA, and St Georg tibial components compared. Acta Orthop Scand 1992; 63: 247–255.
- 36 Hood R, Wright T, Burstein A. Retrieval analysis of total knee prostheses: A method and its application to 48 total condylar prostheses. J Biomed Mater Res 1983; 17: 829–842.
- 37 Furmanski J, Kraay MJ, Rimnac CM. Crack initiation in retrieved cross-linked highly cross-linked ultrahigh-molecular-weight polyethylene acetabular liners. J Arthroplasty Elsevier Inc; 2011; 26: 796–801.
- 38 Kurtz SM, Ochoa JA, White CV, Srivastav S, Cournoyer J. Backside nonconformity and locking restraints affect liner/shell load transfer mechanisms and relative motion in modular acetabular components for total hip replacement. J Biomech 1998; 31: 431–437.
- 39 Williams S, Butterfield M, Stewart T, Ingham E, Stone M, Fisher J. Wear and deformation of ceramic-on-polyethylene total hip replacements with joint laxity and swing phase microseparation. Proc Inst Mech Eng H 2003; 217: 147–153.
- 40 Williams S, Leslie I, Isaac G, Jin Z, Ingham E, Fisher J. Tribology and wear of metal-on-metal hip prostheses: Influence of cup angle and head position. J Bone Joint Surg Am 2008; 90 Suppl 3: 111–117.
- 41 Hua X, Li J, Wang L, Jin Z, Wilcox R, Fisher J. Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions. J Biomech Elsevier; 2014; 47: 3303–3309.
- 42 Galvin AL, Tipper JL, Jennings LM, Stone MH, Jin ZM, Ingham E, Fisher I. Wear and biological activity of highly crosslinked polyethylene in the hip under low serum protein concentrations. Proc Inst Mech Eng H 2007; 221: 1–10.
- 43 Al-hajjar M, Lancaster-jones OOD, Williams S, Jennings LM. Different levels of rotational and translational surgical mal-positioning affects the occurrence and severity of edge loading and wear in total hip replacements. Orthop Res Soc Annu Meet 2015.