radiation damage
Radiation damage in protein crystals examined under various conditions by different methods
aLaboratory of Molecular Biophysics, Department of Biochemistry, Oxford University,
South Parks Road, Oxford OX1 3QU, UK, and bDiamond Light Source Ltd, Diamond House, Harwell Science and Innovation Campus, Didcot,
Oxfordshire OX11 0DE, UK
*Correspondence e-mail: [email protected], [email protected]
Investigation of radiation damage in protein crystals has progressed in several directions over the past couple of years. There have been improvements in the basic procedures such as calibration of the incident
and calculation of the dose likely to be deposited in a crystal of known size and composition with this intensity. There has been increased emphasis on using additional techniques such as optical, Raman or to complement X-ray diffraction. Apparent discrepancies between the results of different techniques can be explained by the fact that they are sensitive to different length scales or to changes in the electronic state rather than to movement of atoms. Investigations have been carried out at room temperature as well as cryo-temperatures and, in both cases, with the introduction of potential scavenger molecules. These and other studies are leading to an overall description of the changes which can occur when a protein crystal is irradiated with X-rays at both cryo- and room temperatures. Results from crystallographic and spectroscopic radiation-damage experiments can be reconciled with other studies in the field of radiation physics and chemistry.Keywords: radiation damage; protein crystals; scavengers; microspectroscopy.
There have been significant advances in the understanding of radiation damage in protein
crystallography over the past few years, with damage to protein crystals at room temperature
as well as cryo-temperature being investigated. A variety of techniques to complement
X-ray diffraction have been used, and studies with added scavengers to potentially
mitigate radiation damage have been carried out. Those scientists involved in these
studies are familiar with the detailed mechanisms and consequences of radiation damage.
However, the majority of crystallographers require easily understood guidelines to
carry out a successful ) is written to fulfil this need. It is also a useful review of radiation damage studies
and includes several interesting suggestions which would benefit from further detailed
investigation. It therefore merits reading by both non-experts and experts alike.
The most relevant metric against which to measure damage is the absorbed dose (energy
per unit mass) deposited in the sample. This depends on the properties of the incident
beam and the energy deposited by this beam in the specimen, which in turn is determined
by the i.e. its constituent atoms. Apparently simple terms such as the incident beam `flux' or
`intensity' are not consistently defined across scientific disciplines. The potential
confusion can only be avoided by careful definition and consistent use of the various
terms, as emphasized in the article by Holton (2009). The accumulated dose received by a sample depends on the fluence (photons mm−2) received by the relevant portion of it during the X-ray exposure. The fluence will
depend on the (photons s−1 mm−2) and the exposure time. The is likewise determined from the beam size and (photons s−1). Unfortunately, the and beam sizes documented for various protein crystallography beamlines throughout
the world are not as reliable as they need to be. Tabulated values can be out of date,
calculated rather than measured, or simply not available. A reliable and convenient
means of measuring the using pin diodes is given in this issue by Owen, Holton et al. (2009
). Although the pin diodes interrupt the beam, they can be used to measure the beam
before and after X-ray data collection and/or to calibrate the ionization chambers
often used during the X-ray exposures. For this to be an effective strategy, values
from the ionization chambers should be routinely recorded in the diffraction image
headers, so they can then easily be converted into The beam size itself is also not simple to define unless a top hat profile, combined
with accurately measured apertures, is available. A Gaussian beam (for example) will
deposit energy non-uniformly into the crystal, causing for instance differential cell
expansion and varying degrees of specific structural damage through the sample. This
phenomenon exacerbates the problems caused by radiation damage since different parts
of the crystal are being affected by varying amounts. Finally, an estimate of the
dose deposited by the incident beam in the crystal is required. The latest developments
of the program RADDOSE (Murray et al., 2004
) to calculate this conveniently for macromolecular crystallography are described
in this issue by Paithankar et al. (2009
).
Spectroscopic methods are increasingly being applied to the study of radiation damage
in protein crystals, including combinations of X-ray diffraction, e.g. Hough et al., 2008). In this issue, papers by McGeehan et al. (2009
) and Owen, Pearson et al. (2009
) describe experimental arrangements which enable routine measurements of this kind,
together with (resonance) Raman spectroscopy. can be used to measure metal ligand distances to an accuracy of 0.02 Å and is therefore
sensitive to small movements of atoms surrounding a metal atom. in the XANES region is a sensitive measure of the of the absorbing atom (e.g. a metal atom) and it is found that metal atoms can be reduced by very low absorbed
doses of X-rays [e.g. approximately 3 MGy to reduce 50% of MnII centres in photosystem II (Yano et al., 2005
)] presumably because of their high Fast changes also occur in the redox state at other centres, and these changes can
be monitored by UV or visible spectroscopy. IR and Raman spectroscopy are sensitive
to bond breakage and formation, ligand binding and conformational change. Finally,
X-ray diffraction is only sensitive to larger movements of atoms or groups of atoms
and is a technique which monitors the structure averaged over both time and many unit
cells. Apparent discrepancies when monitoring radiation damage with different techniques
can be explained by the fact that the changes are being observed on different length
scales or due to alterations in the electronic state. Many of the movements originating
from radiation damage and in principle observable by X-ray diffraction are largely
suppressed at cryo-temperatures. This can lead to the false conclusion that the relevant
atoms are free of changes in The same consideration provides the reason for the success of cryo-methods in `suppressing'
radiation damage in protein crystals.
The course of radiation damage in protein crystallography will be different for the various temperatures (e.g. room temperature, 100 K, 20 K) which are routinely employed, and will depend on the presence or absence of scavenger molecules. A description of the various processes could include the following events.
The primary event is the creation of a photoelectron, created when a photon of around
12 keV is absorbed by a light atom. The photoelectron has a path length of a few micrometres
and, for small crystals, can escape, thus causing less damage than if all the energy
was deposited in the crystal (Nave & Hill, 2005; Cowan & Nave, 2008
). While in the crystal, the photoelectron scatters inelastically off surrounding
atoms creating several hundred and positively charged centres (O'Neill et al., 2002
). These are mobile, even at 100 K (Jones et al., 1987
). They will be attracted preferentially to sites of high causing, for example, reduction at metal centres and more general changes in oxidation
states of atoms, adding to specifically electron deficient functional groups throughout
the protein. Many of these changes can be observed by a variety of spectroscopic techniques
[e.g. UV/Vis spectroscopy (Beitlich et al., 2007
; McGeehan et al., 2009
), Raman (Carpentier et al., 2007
), (Corbett et al., 2007
) and EPR (Utschig et al., 2008
)]. When trapped in water, the hydrated (or more generally solvated) electrons give
a characteristic broad optical (Ershov & Pikaev, 1968
). This spectrum of the hydrated electron builds up during the X-ray exposure but
undergoes partial decay (McGeehan et al., 2009
) when the beam is switched off as some electrons recombine (e.g. with the various positively charged holes).
Fisher & Devlin (1995) have previously investigated the mobility of protons at cryo-temperatures. For both
electrons and protons, tunnelling mechanisms are a possible way of surmounting any
energy barriers which cannot easily be overcome by thermal vibrations at cryo-temperatures.
Tunnelling of protons has been implicated in the mechanism of some enzyme reactions
(Masgrau et al., 2006
).
In this issue, Meents et al. (2009) postulate that changes in bond lengths observed in a peptide are due to hydrogen
abstraction from the amino acids. These hydrogen atoms could be abstracted for example
by radicals produced nearby such as hydroxyls or other hydrogen atoms, or by dissociative
electron capture.
Large movements of other atoms would be suppressed at 100 K because the amorphous
solvent present in protein crystals at cryo-temperatures is a glass (i.e. has the structure of a liquid but with rigidly bound atoms). However, local flexibility
will be present even at 100 K, so small movements could occur. Such movements are
a pre-requisite for identifying damage by techniques such as X-ray diffraction and
(which is sensitive to smaller movements of the atoms than is X-ray diffraction).
The rather unpredictable and ill-defined nature of `flexibility' means that it is
difficult to develop a correlation between the environment of a residue (e.g. exposed to solvent or buried) and the susceptibility to damage (if defined as observable
by X-ray diffraction). This is an area which remains to be explored more thoroughly.
Recent results from acetylcholinesterase crystals irradiated at 100 K and 150 K have
shown this to be a fertile area for obtaining dynamic information on proteins which
can elucidate biological function (Colletier et al., 2008).
In some cases the energy barriers could be too high to be surmounted at reduced temperatures
(e.g. 20 K), thereby demonstrating an apparent decrease in radiation damage at these lower
temperatures. There is evidence from both X-ray diffraction and (Yano et al., 2005; Grabolle et al., 2006
; Corbett et al., 2007
; Chinte et al., 2007
; Meents et al., 2007
) of less radiation damage between 7 and 40 K compared with 100 K. In principle, a
metal atom could be reduced with minimum movement of the surrounding atoms and it
might not be expected that additional protection would be conferred at 40 K compared
with 100 K. However, significant protection (a factor of 30) has been observed as
monitored by XANES measurements (Corbett et al., 2007
). The explanation given by Corbett et al. (2007
) is that `the electrons generated by X-ray radiolysis are randomly distributed with
respect to a metal site and therefore only a small subset would likely be optimized
for an athermal reaction'.
The introduction of scavengers for the more mobile species (electrons and protons)
could be expected to reduce radiation damage at cryo-temperatures and there is some
evidence of this for both specific damage and non-specific damage (Kauffmann et al., 2006; Murray & Garman, 2002
; Southworth-Davies & Garman, 2007
; Holton, 2007
; Borek et al., 2007
). As described above, UV/Vis spectroscopy is a sensitive probe of metal oxidation
states. In work reported in this issue (Macedo et al., 2009
), this has been used to search for scavengers which might slow the rate of reduction
in metalloproteins. One candidate was found to be effective, but still did not allow
a complete diffraction data set to be collected from the oxidized form of the protein.
The use of scavenger molecules is also attractive at room temperature as they have
the potential to interact with many more mobile species than is the case at cryo-temperature.
Recent studies at room temperature have yielded some promising results, with significant
reductions in the intensity loss in the presence of two scavengers. Interestingly
there appears to be a linear decay of the intensity with dose in the presence of scavengers
as opposed to the normally observed et al., 2009).
The non-specific damage results from increasing ionization and radical formation at
a multitude of sites rather than the identifiable vulnerable sites which are damaged
at an early stage. The resultant increase of disorder owing to the movement of many
atoms leads to a loss of resolution with increasing dose. This can be referred to
as global damage. For the global damage effects, there is much less variation in the
susceptibility of protein crystals at 100 K compared with crystals at room temperature.
One of the issues still to be resolved is whether the dose required to reduce the
intensity of the reflections (e.g. to half their initial value) follows a linear relationship with resolution (Howells
et al., 2005) or is best described by a linear increase in B factor with dose [as modelled in the program BEST (Bourenkov & Popov, 2006
)]. This is discussed by Holton (2009
). It is noted that the analysis of Howells covered a much greater resolution range
than that assumed in the program BEST so there could be some sampling issues which explain this apparent discrepancy. Although
both models imply that a single number to define the safe dose limit which would apply
at all resolutions cannot be specified, the guideline of a 30 MGy dose limit given
by Owen et al. (2006
) is nevertheless useful.
Many of the events described for crystals at cryo-temperatures will also take place
at room temperature, but they are normally overwhelmed by the large movements of damaged
and reactive species which can occur during and after data collection at room temperature.
Damage (if defined as a loss in resolution during a diffraction experiment) under
these conditions is much harder to predict and accurate intensity monitoring combined
with programs like RADDOSE and BEST becomes less useful. Following the observation of an inverse dose rate effect at
low dose rates of 6–10 Gy s−1 (Southworth-Davies et al., 2007), it could be anticipated that there would be an optimum dose rate at room temperature.
At very high dose rate, small temperature rises in the uncooled specimen could occur,
hastening the diffusion of damaging species. As the diffusion still occurs even if
the beam is absent, too long a data collection time with a weaker beam could result
in increased damage and a lower recombination rate of mobile and reactive species
owing to their lower density. An additional factor, probably to which the inverse
dose rate effect is attributable, is the timescale required for recombination of damaged
species: a beneficial reaction to set against the adverse ones.
The above description of radiation damage is an attempt by the authors to summarize results obtained by many investigators. Some parts of the description are well established, others much less so.
Acknowledgements
We thank Martin Weik and Ian Carmichael for providing useful comments on this manuscript. We also gratefully acknowledge the local organisational skills of Clemens Schulze-Briese and the generosity of the Swiss Light Source for hosting the Fifth International Workshop on X-ray Damage to Crystalline Biological Samples, during which most of this work was presented, and the funding bodies which made it possible through contributing to the cost: the EU programs BioXhit and MAX-INF 2, Dectris Ltd, the Paul Scherrer Institut and from I3 IA-SFS. We hope to hold the sixth such workshop during 2010, hosted at the SSRL.
References
Barker, A. I., Southworth-Davies, R. J., Paithankar, K. S., Carmichael, I. & Garman,
E. F. (2009). J. Synchrotron Rad. 16, 205–216. Web of Science CrossRef CAS IUCr Journals Google Scholar
Beitlich, T., Kühnel, K., Schulze-Briese, C., Shoeman, R. L. & Schlichting, I. (2007).
J. Synchrotron Rad. 14, 11–23. Web of Science CrossRef CAS IUCr Journals Google Scholar
Borek, D., Ginell, S. L., Cymborowski, M., Minor, W. & Otwinowski, Z. (2007). J. Synchrotron Rad. 14, 24–33. Web of Science CrossRef CAS IUCr Journals Google Scholar
Bourenkov, G. P. & Popov, A. N. (2006). Acta Cryst. D62, 58–64. Web of Science CrossRef CAS IUCr Journals Google Scholar
Carpentier, P., Royant, A., Ohana, J. & Bourgeois, D. (2007). J. Appl. Cryst. 40, 1113–1122. Web of Science CrossRef CAS IUCr Journals Google Scholar
Chinte, U., Shah, B., Chen, Y.-S., Pinkerton, A. A., Schall, C. A. & Hanson, B. L.
(2007). Acta Cryst. D63, 486–492. Web of Science CrossRef CAS IUCr Journals Google Scholar
Colletier, J.-P., Bourgeois, D., Sanson, B., Fournier, D., Sussman, J. L., Silman,
I. & Weik, M. (2008). Proc. Natl Acad. Sci. USA, 105, 11742–11747. Web of Science CrossRef PubMed CAS Google Scholar
Corbett, M. C., Latimer, M. J., Poulos, T. L., Sevrioukova, I. F., Hodgson, K. O.
& Hedman, B. (2007). Acta Cryst. D63, 951–960. Web of Science CrossRef CAS IUCr Journals Google Scholar
Cowan, J. A. & Nave, C. (2008). J. Synchrotron Rad. 15, 458–462. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ershov, B. G. & Pikaev, A. K. (1968). Advances in Chemistry Series, Vol. 81, edited by R. F. Gould, pp. 1–23. Washington: American Chemical Society.
Google Scholar
Fisher, M. & Devlin, J. P. (1995). J. Phys. Chem. 99, 11584–11590. CrossRef CAS Web of Science Google Scholar
Grabolle, M., Haumann, M., Müller, C., Liebisch, P. & Dau, H. (2006). J. Biol. Chem. 281, 4580–4588. Web of Science CrossRef PubMed CAS Google Scholar
Holton, J. M. (2007). J. Synchrotron Rad. 14, 51–72. Web of Science CrossRef CAS IUCr Journals Google Scholar
Holton, J. M. (2009). J. Synchrotron Rad. 16, 133–142. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hough, M. A., Antonyuk, S. V., Strange, R. W., Eady, R. R. & Hasnain, S. S. (2008).
J. Mol. Biol. 378, 353–361. Web of Science CrossRef PubMed CAS Google Scholar
Howells, M. R., Beetz, T., Chapman, H. N., Chi, C., Holton, J. M., Jacobsen, C. J.,
Kirz, J., Lima, E., Marchesini, S., Miao, H., Sayre, D., Shapiro, D. A. & Spence,
J. C. H. (2005). J. Electron. Spectrosc. Relat. Phenom. In the press. (Preprint at https://arxiv.org/abs/physics/0502059.) Google Scholar
Jones, G. D. D., Lea, J. S., Symons, M. C. R. & Taiwo, F. A. (1987). Nature (London), 330, 772–773. CrossRef CAS PubMed Web of Science Google Scholar
Kauffmann, B., Weiss, M. S., Lamzin, V. S. & Schmidt, A. (2006). Structure, 14, 1099–1105. Web of Science CrossRef PubMed CAS Google Scholar
Macedo, S., Pechlaner, M., Schmid, W., Weik, M., Sato, K., Dennison, C. & Djinović-Carugo,
K. (2009). J. Synchrotron Rad. 16, 191–204. Web of Science CrossRef CAS IUCr Journals Google Scholar
Masgrau, L., Roujeinikova, A., Johannissen, L. O., Hothi, P., Basran, J., Ranaghan,
K. E., Mulholland, A. J., Sutcliffe, M. J., Scrutton, N. S. & Leys, D. (2006). Science, 312, 237–241. Web of Science CrossRef PubMed CAS Google Scholar
McGeehan, J., Ravelli, R. B. G., Murray, J. W., Owen, R. L., Cipriani, F., McSweeney,
S., Weik, M. & Garman, E. F. (2009). J. Synchrotron Rad. 16, 163–172. Web of Science CrossRef CAS IUCr Journals Google Scholar
Meents, A., Dittrich, B. & Gutmann, S. (2009). J. Synchrotron Rad. 16, 183–190. Web of Science CrossRef CAS IUCr Journals Google Scholar
Meents, A., Wagner, A., Schneider, R., Pradervand, C., Pohl, E. & Schulze-Briese,
C. (2007). Acta Cryst. D63, 302–309. Web of Science CrossRef CAS IUCr Journals Google Scholar
Murray, J. & Garman, E. (2002). J. Synchrotron Rad. 9, 347–354. Web of Science CrossRef CAS IUCr Journals Google Scholar
Murray, J. W., Garman, E. F. & Ravelli, R. B. G. (2004). J. Appl. Cryst. 37, 513–522. Web of Science CrossRef CAS IUCr Journals Google Scholar
Nave, C. & Hill, M. A. (2005). J. Synchrotron Rad. 12, 299–303. Web of Science CrossRef CAS IUCr Journals Google Scholar
O'Neill, P., Stevens, D. L. & Garman, E. (2002). J. Synchrotron Rad. 9, 329–332. Web of Science CrossRef CAS IUCr Journals Google Scholar
Owen, R. L., Holton, J. M., Schulze-Briese, C. & Garman, E. F. (2009). J. Synchrotron Rad. 16, 143–151. Web of Science CrossRef CAS IUCr Journals Google Scholar
Owen, R. L., Pearson, A. R., Meents, A., Boehler, P., Thominet, V. & Schulze-Briese,
C. (2009). J. Synchrotron Rad. 16, 173–182. Web of Science CrossRef CAS IUCr Journals Google Scholar
Owen, R. L., Rudino-Pinera, E. & Garman, E. F. (2006). Proc. Natl. Acad. Sci. USA, 103, 4912–4917. Web of Science CrossRef PubMed CAS Google Scholar
Paithankar, K. S., Owen, R. L. & Garman, E. F. (2009). J. Synchrotron Rad. 16, 152–162. Web of Science CrossRef CAS IUCr Journals Google Scholar
Southworth-Davies, R. J. & Garman, E. F. (2007). J. Synchrotron Rad. 14, 73–83. Web of Science CrossRef CAS IUCr Journals Google Scholar
Southworth-Davies, R. J., Medina, M. A., Carmichael, I. & Garman, E. F. (2007). Structure, 15, 1531–1541. Web of Science CrossRef PubMed CAS Google Scholar
Utschig, L. M., Chemerisov, S. D., Tiede, D. M. & Poluektov, O. G. (2008). Biochemistry, 47, 9251–9257. Web of Science CrossRef PubMed CAS Google Scholar
Yano, J., Kern, J., Irrgang, K. D., Latimer, M. J., Bergmann, U., Glatzel, P., Pushkar,
Y., Biesiadka, J., Loll, B., Sauer, K., Messinger, J., Zouni, A. & Yachandra, V. K.
(2005). Proc. Natl Acad. Sci. USA, 102, 12047–12052. Web of Science CrossRef PubMed CAS Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.