Determination of Left Ventricular Mass by Real-Time Three-Dimensional Echocardiography: In Vitro Validation

Shin Takuma M.D.

Shin Takuma M.D.

Division of Cardiology, Department of Medicine, Columbia-Presbyterian Medical Center, New York, New York

Search for more papers by this author
Santos E. Cabreriza

Santos E. Cabreriza

Division of Cardiology, Department of Medicine, Columbia-Presbyterian Medical Center, New York, New York

Search for more papers by this author
Robert Sciacca Eng., Sc.D.

Robert Sciacca Eng., Sc.D.

Division of Cardiology, Department of Medicine, Columbia-Presbyterian Medical Center, New York, New York

Search for more papers by this author
Marco R. Di Tullio M.D.

Marco R. Di Tullio M.D.

Division of Cardiology, Department of Medicine, Columbia-Presbyterian Medical Center, New York, New York

Search for more papers by this author
Henry M. Spotnitz M.D.

Henry M. Spotnitz M.D.

Division of Cardiology, Department of Medicine, Columbia-Presbyterian Medical Center, New York, New York

Search for more papers by this author
Shunichi Homma M.D.

Shunichi Homma M.D.

Division of Cardiology, Department of Medicine, Columbia-Presbyterian Medical Center, New York, New York

Search for more papers by this author
First published: 28 July 2003
Citations: 8
Address for correspondence and reprint requests: Shunichi Homma, M.D., Columbia-Presbyterian Medical Center, PH 3-342, 630 West 168th Street, New York, NY 10032. Fax: 212-305-9049; E-mail: [email protected]

Abstract

Twenty-one explanted fixed hearts (14 dogs and 7 pigs) were examined to validate newly developed real-time three-dimensional (RT3D) echocardiography for measurement of left ventricular (LV) mass in vitro and to compare its accuracy and variability with those of conventional echocardiographic measurements. There was an excellent correlation and high degree of agreement for the determination of LV mass between RT3D echocardiography and true mass measurement (r = 0.98; standard error of the estimate [SEE]= 7.3 g; absolute difference [AD]= 2.8 g; y = 1.00 x −4.0, interobserver variability; 5.0%). The conventional echocardiographic methods yielded weaker correlations, larger standard errors, and interobserver variability (area-length method: r = 0.90; SEE = 13.3 g; AD = 13.2 g; 13.3%/ truncated ellipsoid method: r = 0.91; SEE = 14.7 g; AD = 10.5 g; 7.9%/ M-mode: r = 0.91; SEE = 16.2 g; AD = 9.4 g; 15.3%). Determination of LV mass by RT3D echocardiography has a high degree of accuracy and is superior to conventional one- and two-dimensional echocardiographic methods.