Phase-derivative analysis in MR angiography: Reduced Venc dependency and improved vessel wall detection in laminar and disturbed flow
Corresponding Author
Romhild M. Hoogeveen MSc
Imaging Center, University Hospital Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands
Imaging Center, University Hospital Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The NetherlandsSearch for more papers by this authorChris J. G. Bakker PhD
Imaging Center, University Hospital Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands
Search for more papers by this authorMax A. Viergever PhD
Imaging Center, University Hospital Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands
Search for more papers by this authorCorresponding Author
Romhild M. Hoogeveen MSc
Imaging Center, University Hospital Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands
Imaging Center, University Hospital Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The NetherlandsSearch for more papers by this authorChris J. G. Bakker PhD
Imaging Center, University Hospital Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands
Search for more papers by this authorMax A. Viergever PhD
Imaging Center, University Hospital Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands
Search for more papers by this authorAbstract
A problem of current MRA techniques is the inability to accurately depict the vascular anatomy, particularly in areas of disturbed flow. Various reasons, such as intravoxel phase dispersion, saturation, temporal variations, and maximum intensity projection (MIP) nonlinearity, cause a wrong delineation of vessel boundaries. A phase contrast (PC)-based postprocessing operation, the phase derivative (PhD), is introduced to detect phase fluctuations indicating flow. Two-dimensional and three-dimensional angiographic reconstruction algorithms are presented. Mathematical formulas are derived to predict the effect of sampling to flow profiles and the effect on the PhD of these profiles. Numerical, phantom, and preliminary in vivo experiments demonstrate that PhD images do not suffer from phase wraps and allow a velocity dynamic range extension only limited by a differential phase change. It is also shown that PhD MIPs produce higher signal-to-noise ratios than conventional PC angiograms and give a better impression of the anatomy of (stenotic) vessels and of their diameters for both laminar and disturbed flow.
References
- 1 Evans AJ, Richardson DB, Tien R, et al. Poststenotic signal loss in MR angiography: effects of echo time, flow compensation and fractional echo. AJNR Am J Neuroradiol 1993; 14: 721–729.
- 2 Gatenby JC, McCauley TR, Gore JC Mechanisms of signal loss in magnetic resonance imaging of stenoses. Med Phys 1992; 20: 1049–1057.
- 3 Oshinski JN, Du DN, Pettigrew RI Turbulent fluctuation velocity: the most significant determinant of signal loss in stenotic vessels. Magn Reson Med 1995; 33: 193–199.
- 4 Urchuk SN, Plewes DB Mechanisms of flow-induced signal loss in MR angiography. J Magn Reson Imaging 1992; 2: 453–462.
- 5 Wolf RL, Richardson DB, LaPlante CC, Huston J III, Riederer SJ, Ehman RL Blood flow imaging through detection of temporal variations in magnetization. Magn Reson Imaging 1992; 185: 559–567.
- 6 Tsuruda J, Saloner D, Norman D Artifacts associated with MR neuroangiography. AJNR Am J Neuroradiol 1992; 13: 1411–1422.
- 7 Atlas SW, Listerud J, Chung W, Flamm ES Intracranial aneurysms: depiction on MR angiograms with a multifeature-extraction, ray-tracing postprocessing algorithm. Radiology 1994; 192: 129–139.
- 8 Anderson CM, Saloner D, Tsuruda JS, Shapeero LG, Lee RE Artifacts in maximum-intensity-projection display of MR angiograms. AJR: Am J Roentgenol 1990; 154: 623–629.
- 9 Burkart DJ, Felmlee JP, Johnson CD, Wolf RL, Weaver AL, Ehman RL Cine phase-contrast MR flow measurements: improved precision using an automated method of vessel detection. J Comput Assist Tomogr 1994; 18: 469–475.
- 10 Bowen BC, Quencer RM, Margosian P, Pattany PM MR angiography of occlusive disease of the arteries in the head and neck: current concepts. AJR Am J Roentgenol 1994; 162: 9–18.
- 11 Mattle HP, Kent KG, Edelman RR, Atkinson DJ, Skillman JJ Evaluation of the extracranial carotid arteries: correlation of magnetic resonance angiography, duplex ultrasonography, and conventional angiography. J Vasc Surg 1991; 13: 838–845.
- 12 Patel MR, Klufas RA, Kim D, Edelman RR, Kent KC MR angiography of the carotid bifurcation: artifacts and limitations. AJR Am J Roentgenol 1994; 162: 1431–1437.
- 13 Lin W, Haacke EM, Smith AS Lumen definition in MR angiography. J Magn Reson Imaging 1991; 1: 327–336.
- 14 Yuan C, Gullberg GT, Parker DL Flow-induced phase effects and compensation technique for slice-selective pulses. Magn Reson Med 1989; 9: 161–176.
- 15 Bryant DJ, Payne JA, Firmin DN, Longmore DB Measurement of flow with nmr imaging using a gradient pulse and phase difference technique. J Comput Assist Tomogr 1984; 8: 588–593.
- 16 Wedeen VJ, Rosen BR, Chesler D, Brady TJ MR velocity imaging by phase display. J Comput Assist Tomogr 1985; 9: 530–536.
- 17 Dijk PV Direct cardiac NMR imaging of heart wall and blood flow velocity. J Comput Assist Tomogr 1984; 8: 429–436.
- 18 Turski PA, Korosec FR Phase contrast angiography. In: Clinical magnetic resonance angiography. CM Anderson, RR Edelman, PA Turski, editors. New York: Raven Press, 1993; 43–77.
- 19 Lee AT, Pike GB, Pelc NJ Three-point phase-contrast velocity measurements with increased velocity-to-noise ratio. Magn Reson Med 1995; 33: 122–126.
- 20 Anderson CM, Lee RE, Levin DL, de la Torre AS, Salonder D Measurement of internal carotid artery stenosis from source MR angiograms. Radiology 1994; 193: 219–226.
- 21 Edelman RR, Mattle HP, Wallner BW, et al. Extracranial carotid arteries: evaluation with “black blood” MR angiography. Radiology 1990; 177: 45–50.
- 22 Edelman RR, Chien D, Kim D Fast selective black blood MR imaging. Radiology 1991; 181: 655–660.
- 23 Chien D, Goldmann A, Edelman RR High-speed black blood imaging of vessel stenosis in the presence of pulsatile flow. J Magn Reson Imaging 1992; 2: 437–441.
- 24 Wasserman BA, Haacke E Carotid plaque formation and its evaluation with angiography, ultrasound and MR angiography. J Magn Reson Imaging 1994; 4: 515–527.
- 25 Hoogeveen RM, Bakker CJG, Viergever MA Improved visualization of vessel lumina using phase gradient images. In: Proceedings of the 12th annual scientific meeting of the Society of Magnetic Resonance in Medicine. New York: Society of Magnetic Resonance in Medicine, 1995; 578.
- 26 Hoogeveen RM, Bakker CJG, Viergever MA Reduced Venc dependency, enhanced SNR, and improved vessel delineation in phase contrast angiograms with phase gradient images. In: Proceedings of the 4th annual scientific meeting of the International Society for Magnetic Resonance in Medicine. New York, International Society for Magnetic Resonance in Medicine, 1996; 244.
- 27 Uyama C, Kita Y, Matssita S Optimal sampling interval and edge detection algorithm for measurement of blood vessel diameter on a cineangiogram. Invest Radiol 1993; 28: 1128–1133.
- 28 Bernstein MA, Ikezaki Y Comparison of phase-difference and complex-difference processing in phase-contrast MR-angiography. J Magn Reson Imaging 1991; 1: 725–729.
- 29 Polzin JA, Alley MT, Korosec FR, Grist TM, Wang Y, Mistretta CA A complex-difference phase-contrast technique for measurements of volume flow rates. J Magn Reson Imaging 1995; 5: 129–137.
- 30 Batten JR, Nerem RM Model study of flow in curved and planar arterial bifurcations. Cardiovasc Res 1982; 16: 178–186.
- 31 Roberts RA, Mullis CT Digital signal processing. MA Reading: Addison-Wesley, 1987.
- 32 Conturo TE, Smith GD Signal-to-noise in phase angle reconstruction: dynamic range extension using phase reference offsets. Magn Reson Med 1990; 15: 420–437.