Volume 100B, Issue 3 pp. 718-725

In vitro and in vivo testing of a novel, hyperelastic thin film nitinol flow diversion stent

C. P. Kealey

Corresponding Author

C. P. Kealey

Department of Surgery, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, California

David Geffen School of Medicine, Department of Surgery, University of California Los Angeles, Los Angeles, CaliforniaSearch for more papers by this author
Y. J. Chun

Y. J. Chun

Department of Industrial Engineering, Swanson School of Engineering, University of Pittsburgh, Pittsburgh, Pennsylvania

Search for more papers by this author
F. E. Viñuela

F. E. Viñuela

Division of Interventional Neuroradiology, Department of Radiology, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, California

Search for more papers by this author
K. P. Mohanchandra

K. P. Mohanchandra

Department of Mechanical and Aerospace Engineering, Henry Samueli School of Engineering, University of California Los Angeles, Los Angeles, California

Search for more papers by this author
G. P. Carman

G. P. Carman

Department of Mechanical and Aerospace Engineering, Henry Samueli School of Engineering, University of California Los Angeles, Los Angeles, California

Search for more papers by this author
F. Viñuela

F. Viñuela

Division of Interventional Neuroradiology, Department of Radiology, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, California

Search for more papers by this author
D. S. Levi

D. S. Levi

Division of Pediatric Cardiology, Department of Pediatrics, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, California

Search for more papers by this author
First published: 28 November 2011
Citations: 10

How to cite this article: Kealey CP, Chun YJ, Viñuela FE, Mohanchandra KP, Carman GP, VIñuela F, Levi DS. 2012. In vitro and in vivo testing of a novel, hyperelastic thin film nitinol flow diversion stent. J Biomed Mater Res Part B 2012:100B:718–725.

Abstract

A flexible, low profile, flow diversion stent could replace endovascular coiling for the treatment of intracranial aneurysms. Micropatterned-thin film nitinol (TFN) is a novel biomaterial with high potential for use in next-generation endovascular devices. Recent advancements in micropatterning have allowed for fabrication of a hyperelastic thin film nitinol (HE-TFN). In this study, the authors describe in vitro and in vivo testing of novel HE-TFN based flow diverting stents. Two types of HE-TFN with expanded pores having long axes of 300 and 500 μm were used to fabricate devices. In vitro examination of the early thrombotic response in whole blood showed a possible mechanism for the device's function, whereby HE-TFN serves as a scaffold for blood product deposition. In vivo testing in swine demonstrated rapid occlusion of model wide-neck aneurysms. Average time to occlusion for the 300-μm device was 10.4 ± 5.5 min. (N = 5) and 68 ± 30 min for the 500-μm device (N = 5). All aneurysms treated with bare metal control stents remained patent after 240 min (N = 3). SEM of acutely harvested devices supported in vitro results, demonstrating that HE-TFN serves as a scaffold for blood product deposition, potentially enhancing its flow-diverting effect. Histopathology of devices after 42 days in vivo demonstrated a healthy neointima and endothelialization of the aneurysm neck region. HE-TFN flow-diverting stents warrant further investigation as a novel treatment for intracranial aneurysms. © 2011 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2012.

The full text of this article hosted at iucr.org is unavailable due to technical difficulties.