Correction for axial dispersion of data obtained by gel permeation chromatography with an on-line viscometer
Abstract
enCorrection to axial dispersion of the results of analyses of polystyrene samples using a gel permeation chromatograph with an on-line viscometer was carried out by a procedure suggested in a preceding paper. Data obtained with each of the detectors of the apparatus were corrected separately, instead of correction of the molecular weight values calculated using data of both detectors. In this case there is no need to use procedures based on the solution of systems of linear equations: any method can be used which has originally been suggested only for correction of the concentration record, which in our case is the method of Pierce and Armonas.
Abstract
deDie Korrektur der an Polystyrenproben mittels eines Gelpermeationschromatographen mit on-line-Viskosimeter erhaltenen Analysenergebnisse wurde unter Verwendung eines Verfahrens durchgeführt, das in einer vorausgegangenen Arbeit beschrieben wurde. Die von den beiden Detektoren gelieferten Werte werden getrennt korrigiert, anstelle einer Korrektur der Molekülmassen unter Verwendung der Angaben beider Detektoren. Damit besteht keine Notwendigkeit, Systeme von linearen Gleichungen zu lösen, man kann vielmehr jede Methode anwenden, die zur Korrektur der Konzentrationskurve geeignet ist. In dieser Arbeit ist es die Methode von Pierce und Armonas.